10
Never disinfect the probe with steam, heating or liquid sterilization
methods, which will cause permanent damage to the probe.
Do not connect or disconnect the probe during real-time scanning; the
system must be in freeze mode or the system is switched off before
connecting or disconnecting the probe.
Repeated disinfection will lead to the safety of the probe and performance
degradation, should regularly check the performance of the probe.
1.8
Electrical Safety
The ultrasound system must use the matching power adapter.
When the system is used in conjunction with a high-frequency device
(such as a surgical electrode), surgical instrument failure or damage to
the probe lens may result in a surgical electrode burned to the patient.
Remove the probe and heart rate display waveforms from the patient
before performing electrical shock defibrillation.
The system uses high frequency signal. Pacemakers interfere with these
signals. Users should be aware of potential hazards and should
immediately turn off the equipment if they are found or suspected of
interfering with the pacemaker operation.
All the input / output terminals of this machine can only be connected with