EasyABPM: Device/Read data  

 

7. Customize, review & print data 

7.1 Customize your standard report 
 

 CardioVisions: Tools/Options/Standard reports/ABPM report 

 

 

EasyABPM: Tools/Settings/Report 

7.2 Review the study and edit data if necessary. 
 
7.3 Create and save or print the report 

 

ABPM-05 Manual Programming 

ABPM-05  if  programmed  manually  can  be  used  by  CardioVisions  software  1.13  or 
later versions. Plans are stored in the inbuilt memory of the device and they cannot 
be  changed.  The  following  three  measurement  plans  can  be  selected  during  the 
programming of the device: 

PLAN A): measurements every 15 minutes at day and every 30 minutes at night.  
PLAN B): periods with 20 minutes at day and 40 minutes at night.  
PLAN C):30-minute periods independent of day or night time. 
 
Other settings are the same in all the three plans: undecided cuff size, 300 mmHg 
pressure  limit,  LCD  display  enabled,  manual  day/night  shift  disabled.  Daytime 
starts at 6:00, while nighttime starts at 22:00, special session is disabled. Patient 
data can be selected or created later in CardioVisions database. 
 

How to program manually 

 

Press  and  hold  the  Start  and  the  Day/night  buttons 
simultaneously.  Measurement  frequency  of  the 
measurement plans will be displayed after 10 seconds 
for 3-3 seconds.  
 
To choose the measurement plan, release the buttons 
when the specific plan is displayed. You will hear two 
beeps  and  the  LCD  will  display  four  blinking  „o” 
letters,  which  indicates  that  programming  the  device 
is  in  progress.  After  a  successful  programming  you 
can hear 5 beeps, and the selected plan can be seen 
again. If programming fails for some reason, the E90 
error code will be displayed on the LCD. 

In  case  of  manual  programming  there  is  no  time  setting.  If  the  time  setting  is 

Содержание ABPM-05

Страница 1: ...User Guide ABPM 05 BlueBP 05 Ambulatory Blood Pressure Monitors...

Страница 2: ...11 7 Customize review print data 12 ABPM 05 manual programming 12 Patient information 13 Cuffs 14 Dimensions 14 Application 15 Batteries 16 Safety concerns 17 Cleaning protection 18 Maintenance 19 Dis...

Страница 3: ...surge for reliable hypertension management and control Both ambulatory blood pressure monitors incorporate an algorithm validated to BHS British Hypertension Society and AAMI Association for the Advan...

Страница 4: ...tion during measurement current pressure is displayed mmHg The device is switched off Systolic value of just completed measurement mmHg The blood pressure measurement is cancelled by pressing a button...

Страница 5: ...Heart symbol blinking measurement in progress kPa Diastolic value of a just completed measurement kPa Buttons The monitor has 3 buttons start event and day night Any blood pressure measurement can be...

Страница 6: ...event marker in a diary Day night button functions Marking the time of sleeping and awakening If the day night shift function is disabled during programming the patient can press the day night button...

Страница 7: ...k and basic solutions 2 Setup the monitor 2 1 Install 2 AA batteries into the monitor Accumulator voltage should be over 2 5V To check press the Start button for at least 5 seconds 2 2 Setup communica...

Страница 8: ...unit converts optical signals to electric ones and vice versa The twin optical cable transfers optical signals between the interface unit and the recorder The cable is flexible but it is sensitive to...

Страница 9: ...PM 05 select the type of your interface unit select the required port or simply click on Test EasyABPM Device Read data BlueBP 05 If you want to use a BlueBP 05 with Bluetooth under Windows XP SP 2 fo...

Страница 10: ...corder in case of BlueBP 05 is BB5_xxxx where xxxx is the last four digits of the serial number found on the back of the recorder After selecting the appropriate recorder the OK button becomes active...

Страница 11: ...atient a sample is available on the installation CD and inform the patient about the goal and the expected results of the monitoring and about the use of the monitor 6 Retrieve data 6 1 Remove the uni...

Страница 12: ...ght time Other settings are the same in all the three plans undecided cuff size 300 mmHg pressure limit LCD display enabled manual day night shift disabled Daytime starts at 6 00 while nighttime start...

Страница 13: ...minute day night intervals 1 measurement plan 15 30 minute day night intervals Programming is in progress blinking signal Patient Information Wearing a thin shirt under the cuff is recommended It does...

Страница 14: ...Event button the patient can mark events e g taking medication etc The time of going to and rising from bed can be marked by the Day night button Should the batteries run down during a monitoring ses...

Страница 15: ...he cuff is placed above the brachial artery if possible Contrary to the usual placement with the tube pointing downwards the advantage is that the patient can wear a loose jacket over the cuff When pr...

Страница 16: ...ult in measurement error and or in certain cases it causes damage to the main recorder unit Batteries Meditech ABPM 05 and BlueBP 05 operate either with two 1 5V AA normal batteries or with two 1 2V A...

Страница 17: ...sion they can be replaced Monitoring will continue and data will not be lost Remove the batteries if the recorder is out of use If you do not use the recorder it is advisable to remove batteries since...

Страница 18: ...ors are not specially protected against spills or ingression of water or other liquids Cleaning the monitor A recommended means of cleaning is to wipe the recorder with a disinfectant cleaning tissue...

Страница 19: ...itech Ltd see relevant topic This warranty does not cover any malfunction or defects arising from improper use the use of inadequate accessories accident theft or use of the device outside operating e...

Страница 20: ...ldren without supervision or children younger than 8 years Though the blood pressure measurement algorithm used in the monitors has been found to function properly on patients with atrial fibrillation...

Страница 21: ...rement method oscillometric blood pressure maximum storage over 600 measurements measurement range blood pressure 30 260 mmHg 4 35 kPa pulse 40 200 beat minute passive accuracy 3mmHg 0 4 kPa or 2 of m...

Страница 22: ...tient did not wear the cuff device error E90 device error The device could not measure due to a hardware error E99 device error The device does not start automatic measurement until the next programmi...

Страница 23: ...accept for repair potentially infectious products or accessories especially pouches and cuffs that might have been in direct contact with the patient and could not be or potentially were not properly...

Страница 24: ...ments Note that any blood pressure recording may be affected by the body position the physiological condition of the patient and other factors No user serviceable parts inside Meditech recorders conta...

Страница 25: ...ected against splash vertically in a max of 15 degrees This protection refers only to BP5 and BB5 SN YYYY nnnnn Serial number The first four digits of the serial number of a recorder show the year of...

Страница 26: ...in domestic establishments and in establishments directly connected to the low voltage power supply network which supplies buildings used for domestic use Harmonic emission IEC61000 3 2 Not applicable...

Страница 27: ...power disclosed by the manufacturer of transmitter W and d is the recommended separation distance m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey1 a should...

Страница 28: ...wer of the transmitter W 150KHz 80MHz d 3 5 V1 P 80MHz 800MHz d 3 5 E1 P 800MHz 2 5GHz d 7 E1 P 0 01 Not applicable 0 12 0 23 0 1 Not applicable 0 38 0 73 1 Not applicable 1 2 2 3 10 Not applicable 3...

Отзывы: