Tissue Flotation Bath 

 

TFB 55

 

 

Instruction Manual 

 

 

 
 
 

IMPORTANT: Read this manual before starting the instrument  

 
 
 

Rev. 00 (04/2010) 

 
 
 
 

© Copyright: 

Duplication of this manual by any means for any purpose without written consent of 

Medite GmbH is prohibited. 

 
 
 
 

 

MEDITE GmbH - Wollenweberstr. 12 - D-31303 Burgdorf - Tel: +49 5136 8884 0  -  Fax: + 49 5136 8884 55  [email protected]   

www.medite.de

 

 

Содержание TFB 55

Страница 1: ...Read this manual before starting the instrument Rev 00 04 2010 Copyright Duplication of this manual by any means for any purpose without written consent of Medite GmbH is prohibited MEDITE GmbH Wollen...

Страница 2: ...E Guidelines indicated below including all changes and addendums to date thereto The above mentioned product has been controlled and meets the following standards and guidelines EU Directive 2004 108...

Страница 3: ...e 1 1 3 Authorized Operator 1 1 4 Safety measures on site 1 2 Transport Installation 2 2 1 Technical data 2 2 2 Operation conditions 2 2 3 Transport Storage 3 2 4 Unpacking 3 2 5 Set up and connection...

Страница 4: ...reasons For the exchange of defective parts only original medite spare parts have to be used The conditions for operation maintenance and service mentioned in this instruction manual have to be strict...

Страница 5: ...system avoids overheating and damage even in case of empty bath Equipment mains switch temperature control with scale in C 2 2 Operating conditions The operation is only allowed in closed rooms labor...

Страница 6: ...whether the equipment is complete The TFB 55 is supplied with one power cord and one instruction manual 2 5 Installation and Connection Before starting the instrument please make sure that your mains...

Страница 7: ...ne by the enclosed rubber ring being rolled over the thermometer from above 2 Fill with distilled or fully desalinated water up to 1 2 cm below the rim of the bath If tap water is used the black inter...

Страница 8: ...y turning the temperature regulator knob Always turn clockwise Pilot lamp lights up and extinguishes when the temperature is achieved Before the required temperature is achieved heating will be switch...

Страница 9: ...damage the black epoxy coating In case of a non functioning of the instrument e g caused by voltage oscillations please check whether the fuses in the back of the instrument are defect Should the def...

Страница 10: ...riod of use please deliver the equipment to an authorized disposal facility The company Medite as manufacturer is also prepared to take back the equipment for disposal against payment for the disposal...

Страница 11: ...Rev 00 04 2010 HE TFB55e_Druck_00 DOC 8 7 Wiring diagram...

Страница 12: ...ES 99 430 Width 360 mm Depth 570 mm Height 300 mm Cooling plate 210 x 260 mm Temperature adjustable from 15 C to 10 C Weight 20 kg Voltage 230V 50Hz 150VA Dispenser Unit TES 99 200 Width 360 mm Depth...