P3V Veterinary Digital Ultrasonic Diagnostic Imaging System User Manual
- 52
-
You must restart the system to validate the change, including
Language
,
Keyboard Sound
,
Cursor Time
. After you perform those presetting, and press
Return
, the system displays a
confirm dialog box to prompt you whether to restart the system.
5.7.4. Presetting Examination
Examination types include LAbd, SAbd, OB_Cannine, OB_Feline, OB_Equine, OB_Bovine,
OB_Ovine, OB_Other, Muc, Ten, Sml, Car, Vas.
Take
Canine Ob
examination presetting for example, in the preset menu, move the cursor to
highlight
Canine Obstetric
and press
Set
to display obstetric examination presetting dialog box.
B/M Parameter Tab
Figure 5-15 Obstetric Presetting -- B/M parameter Tab
Item
Setting
Description
Display
B, 2B, 4B, B/M, M, PW
Set display mode type.
Depth
19 mm ~ 324 mm (C361-2)
Set examine depth.
Focus Num
1/2/3/4
Set the number of focuses.
Scan Density
L/M/H
Set scanning density. Three levels adjustable: low,
middle, high
SRA
0/1
Turn on or off the synthetic receiving aperture (SRA
can be set only when the scan density is set as
“M”).The purpose of SRA is to improve the lateral
resolution
and
reduce
the
image
noise.
Synthesizing two receive apertures will improve the
signal to noise ratio.
Line Persist
0~7
Set image line correlation.
Probe
Display all the probe type
this device supports
Set the probe type to use.