P3V Veterinary Digital Ultrasonic Diagnostic Imaging System User Manual
- 115
-
measured result window.
9.2.2. RV (Right Ventricle Internal Diameter)
1.
In the B-cardiac measurement menu, roll the trackball to highlight
RV
.
2.
Measure RV with distance method.
3.
The result will be displayed in the measured result window.
9.2.3. PA (Pulmonary Artery)
1.
In the B-cardiac measurement menu, roll the trackball to highlight
PA
, and press
Set
to
display a “+” in the image area.
2.
Measure
PA
with distance measurement method.
3.
The result will be displayed in the measured result window.
Other parameters:
If you want to perform other cardiac parameter measurements, please enter B/M mode or M
mode cardiac measurement.
The result of ventricle volume measurement is more exact in two-dimension. You can get the
two-dimension heart image of end diastolic and end systolic exactly and conveniently in the B/M
mode. So we suggest that you do the cardiac measurement and calculation in the B/M mode.
9.3. Cardiac Measurement and Calculation in PW Mode
1.
Press
Exam,
select the exam mode
Cardiac
, and then press
Set
.
2.
In PW mode, Press
Set
to set the sample line, and press
Measure
, the system will enter PW
mode cardiac measurement. The PW mode cardiac measurement menu is shown as below:
Figure 9-9 PW Mode Cardiac Measurement and Calculation Menu
The operations of
Velocity
and
Heart Rate
are the same as those described in section
6.6.3
Generic Measurement in PW mode
a
nd
section 6.6.2 Generic Measurement in M mode.