5.4. Signals
Font + Tone
Meaning
Tall + high frequency
An alarm with high priority is active – immediate action by the user is
necessary.
Small + mean frequency
An alarm with medium or low priority is active – prompt reaction by the
user is needed.
only Tone
Flashing battery symbol with continuous tone and one gas is missing
6.
Cleaning and Maintenance
6.1.
Cleaning
The surface of the device should be disinfected prior to the initial use as well as after each use on a patient. For
this purpose, a 70% isopropyl alcohol solution can be used.
Warning – Cleaning
-
The device should never be sterilized or immersed in liquid solutions.
-
No abrasive materials should be used on the surface.
-
Liquids should not get inside the device.
6.2. Exchange of the Internal Batteries
-
The internal rechargeable battery must be replaced as required or every three years at the latest.
The replacement battery must be a rechargeable battery of the same type (can be ordered from
medin, REF 51095).
Warning
The internal battery may only be replaced by the battery types listed above and only by trained, professional
service personnel and in accordance with the instructions in the service manual.
After each maintenance or repair of the medinSINDI
®
, a complete function test must be performed and passed
before this medinSINDI
®
can be used on a patient once again.
6.3. Repairs
Warning
Repairs may only be made to the medinSINDI
®
by specially trained, professional service personnel in
accordance with the instructions and warnings in the service manual. After each maintenance or repair of the
medinSINDI
®
, a complete function test must be performed and passed before this medinSINDI
®
can be used on
a patient once again.
6.4. Device Modifications
Warning
The medinSINDI
®
may not be modified without permission from the manufacturer and subsequent appropriate
examinations and testing to ensure continued safe use.
6.5. Power Supply Unit (Cleaning, Maintenance, Repairs and Modifications)
Cleaning:
Prior to cleaning, the power supply unit of the medinSINDI
®
must be disconnected from the mains.