4.2.
Display Elements
Figure 5: medinSINDI
®
Monitor
Features of the display:
A.
nCPAP pressure display: analogous graph in cmH
2
O
B.
nCPAP pressure display in cmH
2
O (Fair value)
C.
Oxygen concentration display: as bar chart
D.
Oxygen concentration display in %
E.
Triggering of spontaneous respiration: blinking
F.
Spontaneous respiratory rate display: Breaths per minute
G.
Upper (a) and lower (b) pressure alarm limits
H.
Loudspeaker symbol: to indicate alarm mute setting
I.
Battery charge monitoring display
Up to software version 1.016 only a pressure of a maximum of 10 cmH
2
O can be measured and displayed in
pressure curve (A) and at pressure value (B).
Starting with software version 1.017 values up to 10 cmH
2
O are displayed in pressure curve (A) and a pressure
up to 35 cmH
2
O is displayed at pressure value (B).
Starting with software version 1.021 the mute setting must be activated
Starting with software version 1.022 current CPAP low and CPAP high values change in conjunction
Starting with software version 1.023 larger visual alarm displays, automatic trending stop in the event of an
alarm
Starting with software version 1.024 Update via RS232 Interface
4.3.
Starting Up and Adjusting Settings
Important Note – Start-Up
Training after installation by authorized sales and service partners!
For use by trained professionals only!
After long idle times, oxygen calibration must be performed - see service menu oxygen calibration (page 20).
Ga
A
B
C
D
E
F
Gb
I
H