Performing a Test
66
EasyOne Pro™ / EasyOne Pro™ LAB
3.9 Ending the Test
At the end of the test, push the
spirette™
out of the sensor from below.
Dispose of it and do not reuse. Also, replace the
barriette™
after each
patient. Refer to section
"Hygiene, Cleaning, Maintenance, Disposal"
page 89 for sensor cleaning instructions.
Warning
Risk of Infection—
spirette™
and
DLCO barriette™
/
FRC barriette™
are
intended for single patient use. Use new ones for each
new patient.
After tests on patients with an infection of the respiratory
tract or patients with a suspected infection of the
respiratory tract, it is recommended to clean all parts
(outside of the sensor) that were touched during the test,
before testing a new patient (see
3.10 Retrieving/Printing Stored Tests
EasyOne Pro™
saves all tests. Old measurements can be called up at any
time to
perform a new test with the same patient
add a spirometry trial
add a post-medication test
review or print the results.
Note
It is only possible to add a trial to a previous test that was
performed on the same day (see
Select the patient (see