EasyOne Pro™ / EasyOne Pro™ LAB
5
Introduction
1
Introduction
1.1 CE Marking Information
The product
EasyOne Pro™
bears the CE marking CE-120 (notified body
SGS) indicating its compliance with the provisions of the Council
Directive 93/42/EEC about medical devices and fulfills the essential
requirements of Annex I of this directive.
The flow sensor has been assigned to class IIa as specified in Annex IX of
the Directive 93/42/EEC.
The device fulfills the requirements of standard EN 60601-1 "Medical
electrical equipment, Part 1: General requirements for basic safety and
essential performance" as well as the electromagnetic immunity
requirements of standard EN 60601-1-2 "Electromagnetic compatibility
– Medical electrical equipment".
The radio-interference emitted by
EasyOne Pro™
is within the limits
specified in EN 55011, class B.
The CE marking covers only the accessories listed in the Order
Information chapter.
Magnetic and electrical fields are capable of interfering with the proper
performance of the device. For this reason make sure that all external
devices operated in the vicinity of the equipment comply with the
relevant EMC requirements. X-ray equipment, MRI devices, radio
systems, and cellular telephones are possible sources of interference as
they may emit higher levels of electromagnetic radiation. Keep the
equipment away from these devices and verify its performance before
use.
The device is suitable for continuous operation.
The product fulfills the requiremens of the following standards:
EN ISO 14971
IEC60601-1: 2005
EN 60601-1: 2006
EN 60601-1-2: 2007
EN 60601-1-6: 2007
The country of manufacture is indicated on the device label.