Cardio M Plus 

 

 

 

 

 

Spirometer 

Opration Manual 

 
 
 
 

 

Содержание Cardio M Plus

Страница 1: ...Cardio M Plus Spirometer Opration Manual...

Страница 2: ...ted by Medical ECONET GmbH to be free from defects in material and workmanship for one year Two years in Europe from date of purchase Warranty repair or replacement will be made by medical ECONET GmbH...

Страница 3: ...onnel Product and Purchase Inquiry Medical ECONET GmbH Mainstrasse 6c d D 45768 Marl Germany Te l 49 0 2365 92737 0 Fax 49 0 2365 92437 55 E mail info medical econet com URL http www medical econet co...

Страница 4: ...s no responsibility for any kind of damage or breakdown that is caused by misusage and failure to administer the equipment Warning To inform that it may cause serious injury or death to the patient pr...

Страница 5: ...If there are any problems with the equipment please follow the steps below Contact the medical ECONET GmbH immediately After gathering the model name Serial Number date of purchasing and description...

Страница 6: ...g humidity ranges from 30 to 85 Avoid in the vicinity of Electric heater Avoid placing in an area where there is an excessive humidity rise orventilationproblem Avoid placing in an area where thereisa...

Страница 7: ...d to protect the patient from electrical shock Do not use the equipment in the vicinity of flammable anesthetics and solvents The equipment conforms to Class A according to IEC EN 60601 1 Safety of El...

Страница 8: ...hed off Device part switched on This symbol identifies a safety note Ensure you understand the function of this control before using it Control function is described in he appreciate operation manual...

Страница 9: ...in body 17 2 Use Spirometer 18 1 Enter patient information 19 1 1 Enter ID 20 1 2 Enter NAME 20 1 3 Enter Age 21 1 4 Enter Gender 21 1 5 Enter Height 22 1 6 Enter Weight 22 1 7 Enter Race 22 1 8 Enter...

Страница 10: ...Spirometer Opration Manual Version 6 1 10 1 Install Spirometer Appearance of Spirometer...

Страница 11: ...he main body as shown below Plan of the back of the main body Grounding terminal Part connected to external grounding terminal when there is no protection grounding in a socket Serial cable connection...

Страница 12: ...2 Upper cover lock Part that locks upper cover fixing the mouthpiece 3 Action lamp Lamp showing action status green 4 Connection conductor Conductor connected to serial cable connection hole of serial...

Страница 13: ...cover Upper cover that fixes mouthpiece 2 Upper cover lock Part that locks upper cover fixing the mouthpiece 3 Connection conductor Conductor connected to serial cable connection hole in the back of t...

Страница 14: ...1 Upper cover Upper cover that fixes the mouthpiece 2 Power switch Switch that turns on or off the power 3 Connection conductor Conductor connected to serial cable connection hole in the back of the m...

Страница 15: ...the mouthpiece 2 Upper cover lock Part that locks upper cover fixing the mouthpiece 3 Power switch Switch that turns on or off the power 4 Connection conductor Conductor connected to serial cable conn...

Страница 16: ...de of the spirometer handle to take measurement First press Upper cover lock in the upper left side of spirometer handle and open semicircle upper cover and insert the mouthpiece into a groove made in...

Страница 17: ...above connect the power cable to the spirometer cable in the main body 2 Connect spirometer handle to main body while the power of the main body is off 3 Turn on power switch of spirometer handle 4 Tu...

Страница 18: ...IRO by turning lottery switch and then enter each menu by pressing lottery switch If you choose spirometer in initial menu following initial menu appears on the LCD of the main body Picture1 Spiromete...

Страница 19: ...y switch and keyboard that appears on the LCD After you complete entry or if you press Cancel key while entering information the screen turns to the main screen without saving If you press New key inf...

Страница 20: ...m left top to right bottom Letters or numbers on which the cursor is placed are highlighted on the screen If you press the lottery switch or touch there the selected letters or numbers are entered If...

Страница 21: ...switch gender of the patient appears alternatively as male and female If you turn the lottery switch to clockwise after selecting patient s gender cursor moves to Height entry space Patient Informati...

Страница 22: ...er RACE Enter patient s race The race is registered as Oriental Black and Caucasian If you press the lottery switch in the same way as you did when you entered patient s gender above mentioned races a...

Страница 23: ...ck day indicates the smoking volume of the patient per day Select one of 1 4 1 2 1 2 2more by pressing lottery switch If you turn the switch to clockwise it is placed on Years Enter the period that th...

Страница 24: ...UP LIST 10 CALIB 2 FVC Forced Vital Capacity TEST Select FVC TEST by turning the lottery switch to clockwise from the initial menu and press the lottery switch Once you begin to take measurement messa...

Страница 25: ...th and begins to breathe the machine takes measurement If spirometer handles are not connected or power is off following message appears This message appears as the same in SVC test and MVV test FVC T...

Страница 26: ...possible to avoid panting 5 After that you should signal the machine that you have finished taking the measurement by pressing the stop button on the screen If you have difficulties in taking breaths...

Страница 27: ...reaths shown below is drawn and Volumes Flows Times and Ratios appear in about one and a half seconds Result screen Signal the machine that measurement is finished by pressing Stop button by lottery s...

Страница 28: ...ew menu in beginning the 2 nd Test after completion of the 1 st measurement FVC Test F L s 0 Screen after NEW is clicked If you click Yes here screen begins the second measurement as below Then press...

Страница 29: ...ew the detailed graph on the result FVC measurement result screen View 1 At this status if you click lottery switch once more or press NEXT by touch screen of detailed figures on the result appears FV...

Страница 30: ...reen 1 If would like to save the data choose Yes If not choose No 1 1 In case of choosing Yes 1 The patient information entering window will be opened after a message Enter Patient ID if there is no p...

Страница 31: ...ack to main screen of Spirometer In case of full memory there will be message as follows The message after saving last one SYSTEM MESSAGE Save Succeed Memory is full After all possible data is saved t...

Страница 32: ...POST menu Test should be done after checking FVC menu is changed to FVC POST Saving test result once only for before having medicine 1 Prt Del Info SPR No ID Name Date Type 1234567890123 John 07 11 23...

Страница 33: ...th of before and after having medicine datas Note In case of there is no need to test for difference between before and after having medicine please test FVC at initial menu only Note In case of new p...

Страница 34: ...Spirometer Opration Manual Version 6 1 34 Result of one time test...

Страница 35: ...Spirometer Opration Manual Version 6 1 35 Result of twice test...

Страница 36: ...ore of normal breath it beeps 2 Exhale slowly as much as possible to the end RV level 3 Inhale slowly as much as possible to the end TLC level 4 Return to normal breathing If you breathe in above ment...

Страница 37: ...ished a graph V T showing change in volume of lung is drawn as follows according to passage of time SVC Test 40sec SVC TEST result screen Remained four menus like New Print View Exit are used in the s...

Страница 38: ...ders Unless the patient follows the orders it is not possible to take accurate measurement 1 Begin to take measurement of vital capacity 2 Instruct the patient to breathe as fast and much as possible...

Страница 39: ...he volume is the result of adding amount of inhalation and that of exhalation MVV test 150 L 12sec Screen in measurement Press Stop button after completing measuring If you press Stop button the resul...

Страница 40: ...rs If you choose and press each item Pressure Temperature Humidity Syringe Size by rotary switch or touch number entry space that in times of patient Info appears Enter the responsible value for each...

Страница 41: ...nd syringe amount cc 5 Press start button If the machine is ready for the calibration it beeps 6 Move syringe handle back and forth five times each Make sure to move the handle after you hear the mach...

Страница 42: ...spirometer and the initial menu appears Otherwise the initial menu immediately appears If the touch screen dosen t work properly please enter Touch Screen Set by using rotary key and then please touc...

Страница 43: ...ing ON OFF and Pacemaker signal and Lead fault signal ON OFF are available When selecting by using the rotary key you can change the status of setup and OK and Cancel button will appear When touching...

Страница 44: ...up is completed 3 Unit This is setup of unit when you input patient information as weight and height You could choose between Kg cm or lb in 4 Save After test Save the data window could be disappeared...

Страница 45: ...s not available to use automatic IP allocation function based on windows system You must allocate device s own IP address manually Also DNS setup is not available DB Server means the computer IP addre...

Страница 46: ...Spirometer Opration Manual Version 6 1 46 Hosp Setup Input the doctor s name and hospital name...

Страница 47: ...18 13 MVV 10 1234567890123 John 07 11 29 21 43 ECG SPR Info Del Prt Data management 1 Window and function description Saved data Previous page Next page Current cursor position Number of all data Prev...

Страница 48: ...to PC is being tried SYSTEM MESSAGE DB Connecting Data is being sent SYSTEM MESSAGE Sending Saved Data It indicates the transmission of data is completed SYSTEM MESSAGE Transmission Complete System m...

Страница 49: ...or stain Wipe the cable with a cloth that is wet with warm water 40 C 104 F and with the clinical alcohol once a week Do not soak the machine or the electrocardiogram cable into any liquid or detergen...

Страница 50: ...y the buzzer rings for three times consecutively and the following message is displayed on the screen SYSTEM MESSAGE Battery Low The battery is almost used up Connect the AC power supply before using...

Страница 51: ...11 l Measuring method Differential pressure method Prediction equation Morris Polgar ECCS Quanjer Knudson ITS Sample rate 200 samples sec Flow impedance 0 2 mbar s l at 12 l s Measuring accuracy Compl...

Страница 52: ...chase Two years in Europe Date of Purchase Customer Hospital Address Name Tel Sales Agency Manufacturer Thanks for purchasing our product This product is manufactured and passed through strict quality...

Страница 53: ...1 53 International Sales service Medical ECONET GmbH Mainstrasse 6c d D 45768 Marl Germany Tel 49 0 2365 92437 0 Fax 49 0 235 92437 55 e mail info medical econet com Website www medical econet com Me...