MEDI NUCLEAR Insta/vent PLUS Скачать руководство пользователя страница 1

Step by Step

Instructions

(800) 321-5981, (626) 960-9822

[email protected]

www.medinuclear.com

4610 Littlejohn Street, Baldwin Park, CA 91706

© 2018 Medi/Nuclear® Corporation, Inc. 
All Rights Reserved 

Rev 07/02/18 

Содержание Insta/vent PLUS

Страница 1: ...ep by Step Instructions 800 321 5981 626 960 9822 info medinuclear com www medinuclear com 4610 Littlejohn Street Baldwin Park CA 91706 2018 Medi Nuclear Corporation Inc All Rights Reserved Rev 07 02...

Страница 2: ...llows for the radioaerosol particles of 99mTc DTPA or an equivalent alternative to be immediately available upon inhalation An airflow control device is used instead of one way valves thus eliminating...

Страница 3: ...hrough the nebulizer into a breathing bag which acts as a reservoir to store aerosol that would normally be wasted during exhalation When the next breath is taken the patient immediately inhales the a...

Страница 4: ...bag containing Safety Shield Mouthpiece nose clip and Caution Radioactive Material label Note Retain the re sealable bag for disposal of the Insta Vent Plus after the ventilation study 800 321 5981 62...

Страница 5: ...will be directed into the lead cylinder and into the silver oxygen air pressure port C The HEPA filter will extend through the exhaust opening at the bottom Press down firmly so that the nebulizer is...

Страница 6: ...ter The HEPA filter can be reached through the exhaust opening in the underside of the Insta Vent shield Recheck the manifold to be sure it is seated firmly in place 800 321 5981 626 960 9822 info med...

Страница 7: ...inlet connection on the front of the lnsta Vent lead shield Use only the provided oxygen tube IV 605 Other oxygen tubes may be slightly over sized and could cause oxygen leakage 800 321 5981 626 960 9...

Страница 8: ...y stopper in the top of the Insta Vent manifold and into the nebulizer Prior to closing the Insta Vent shield ensure that the manifold is seated firmly into the silver oxygen air pressure port Close t...

Страница 9: ...ty Shield Mouthpiece cover by pressing in the sides and gently pulling out Place the mouthpiece in the patient s mouth Place the nose clip on the patient s nose unless using a face mask Prior to turni...

Страница 10: ...t of radioactivity is delivered to the lungs Breathing time should be approximately 2 minutes if the assembly recommendations are followed Caution To prevent possible radiation leakage into the enviro...

Страница 11: ...iva into the cover of the mouthpiece Attach the cover to the mouthpiece making sure it seals securely Discard the sealed mouthpiece into the plastic bag Start the patient imaging procedure as soon as...

Страница 12: ...ilter Remove the Insta Vent shield lid and remove the used radioaerosol kit from the shield Disconnect oxygen tubing from the oxygen air source Warning Disconnecting the oxygen line before removing ki...

Страница 13: ...bel and place the kit in a shielded disposal area to allow for radioactive decay Discard decayed waste according to the radioactive waste procedures established by your facility 800 321 5981 626 960 9...

Страница 14: ...Kit TwoTube System 24 cs IV 600PM Insta Vent Radioaerosol Kit with Mask TwoTube System 24 cs IV 600V Insta Vent Ventilator Kit for ventilator dependent patients 6 cs IV 601A Insta Vent Portable Lead S...

Страница 15: ...960 9822 info medinuclear com www medinuclear com We re here to be of assistance 800 321 5981 626 960 9822 info medinuclear com www medinuclear com 4610 Littlejohn Street Baldwin Park CA 91706 2018 Me...