To ensure that the breastpump will provide trouble-free operation for
years to come, it should be examined periodically according to the
following checklist
Checklist:
1 The specification of the fuse employed as well as the power supply
to which the pump is connected must match the data on the pump
specification plate
2 Fit the accessories to the breastpump as described in the
instructions for use (included with the accessory kit) The
breastshield should then be connected to a vacuum gauge
3 Connect the power cord to the main supply and switch on the unit
The pivot arm must operate with a pumping motion
4 The following values must be attained:
Suction rate: Lactina Select
approx 40–60 cycles/min,
mains
approx 50 cycles/min,
battery
Suction strength at MAX setting:
6. Safety Tests
The following safety tests should be performed yearly:
1. Visual check
The breastpump should be examined for damage and dirt on the inside
and outside and repaired or cleaned if necessary
2. Functional test
Every function of the breastpump must be controlled according to the
Instructions for Use and Service Instructions
3. Electrical tests
• Enclosure leakage current: This test should be performed under
normal conditions and under the conditions of Single Fault
Conditions of Single Fault: The interruption of each supply conductor
one at a time
14
Location
0 m above
500 m above
1000 m above 2000 m above
sea level
sea level
sea level
sea level
Factor
1.00
0.94
0.88
0.78
Lactina Select
(+/- 20 mmHg)
255 mmHg
240 mmHg
225 mmHg
199 mmHg