Disassembly and Assembly
Confidential Information of MEDCAPTAIN MEDICAL TECHNOLOGY CO., LTD. 67 / 93
No.
Part Name
Quantity
1
HP-60 pump door hole plug
7
2
PT2.3*6 cross recessed pan head drilling
PT thread tapping screw
7
3
HP-60 cover plate
1
4
HP-60 pump door PCBA
1
5
PT2.3*6 cross recessed pan head drilling
PT thread tapping screw
7
1.
Use tweezers to remove the seven pump door hole plugs (1).
2.
Remove the seven PT2.3*6 cross recessed pan head drilling PT thread tapping screws (2).
3.
Take down the HP-60 cover plate (3).
4.
Remove the seven PT2.3*6 cross recessed pan head drilling PT thread tapping screws (5).
5.
Take out the HP-60 pump door PCBA (4).
1.13
Dismantling the Separate Air Bubble Sensor Module B