Disassembly and Assembly
Confidential Information of MEDCAPTAIN MEDICAL TECHNOLOGY CO., LTD. 61 / 93
No.
Part Name
Quantity
1
HP-60 common consumable middle plate
1
2
PM3*6 screw
4
3
Peristalsis pump assembly
1
1.
Take down the HP-60 sliding block according to the method in section 8.6.
2.
Take down the HP-60 common consumable middle plate (1).
3.
Screw down the four PM3*6 screws (2).
4.
Remove the HP-60 door potentiometer assembly (9121000008), HP-60 motor speed measuring
board’s connection cable (7000000025), and plug of the peristalsis pump motor’s connection
cable.
5.
Take down the peristalsis pump assembly (3).
1.8
Dismantling the Peristalsis Pump Motor