Disassembly and Assembly
Confidential Information of MEDCAPTAIN MEDICAL TECHNOLOGY CO., LTD. 60 / 93
1
HP-60 tube detection PCBA
1
2
PT2.3*6 cross recessed pan head drilling PT
thread tapping screw
2
3
HP-60 anti-free-flow clamp PCBA
4
PT2.3*6 cross recessed pan head drilling PT
thread tapping screw
2
5
HP-60 patient side occlusion pressure
detection PCBA
1
6
PT2.3*6 cross recessed pan head drilling PT
thread tapping screw
2
7
HP-60 patient side occlusion pressure
detection board’s connection FFC
1
1.
Take down the HP-60 rear panel socket assembly and HP-60 AC shield.
2.
Remove the HP-60 patient side occlusion signal connection cable (7008000019).
3.
Screw out the two PT2.3*6 cross recessed pan head drilling PT thread tapping screws (2), and
take down the HP-60 tube detection PCBA (1).
4.
Screw out the two PT2.3*6 cross recessed pan head drilling PT thread tapping screws (4), and
take down the HP-60 anti-free-flow clamp PCBA (3).
5.
Remove the HP-60 patient side occlusion pressure detection board’s connection FFC (7).
6.
Screw out the two PT2.3*6 cross recessed pan head drilling PT thread tapping screws (6), and
take down the HP-60 patient side occlusion pressure detection PCBA (5).
1.7
Dismantling the Peristalsis Pump