Alarms and Troubleshooting
47
9.4
System Faults and Troubleshooting
System Faults and Troubleshooting
Alarm
Message
Alarm Content
Alarm
Level
Alarm Cause
Troubleshooting
Patient Side
OCCL
Patient Side OCCL
High
Real-time pressure in the
tube ≥ occlusion alarm
value configured
1. Check whether the infusion tube is
twisted or extruded and whether the
retaining needle leaks.
2. Check whether the occlusion level
setting is reasonable (an older patient
indicates a higher drug liquid viscosity
and higher infusion rate, and the
occlusion level also needs to be
higher).
3. Calibrate the pressure value of the
consumable.
4.
Calibrate the pressure value of
the pressure sensor when it does
not touch the infusion tube, and
check that the KB value is correct.
(Generally, the K value is
404mV/N±50 and B value is about
500mV±200.
)
5. Check the pressure sensor.
6. Check the drive head board, nut
displacement detection board, main
board, and related connection cable.
ERR 001
System error: motor
drive locked
(motor rotation
speed is abnormal)
High
1. The motor is locked.
2. The motor rotation
speed is too low.
3. The motor rotation
speed is too high.
1. Check the optocoupler and related
connection cable.
2. Check the motor and related
connection cable.
3. Check the speed measurement board
and related connection cable.
4. Check the main board.
ERR 002
System error:
reversed rotation of
motor drive
(motor rotation
direction is
abnormal)
High
The motor rotation
direction is opposite to the
expected direction.
1. Check the optocoupler and related
connection cable.
2. Check the speed measurement board
and related connection cable.
ERR 003
System error:
pressure sensor
failure
(pressure sensor is
faulty)
High
The pressure sensor status
is incorrect.
1. Check the pressure sensor and
related connection cable.
2. Check the drive head board and
related connection cable.
3. Check the main board.
ERR 005
Main chip voltage
3.3V
(CPU voltage is
abnormal)
High
The monitored voltage is
smaller than 1.67V or
greater than 2.25V.
1. Check the main board and related
connection cable (generally, check
whether the C71 voltage on the main
board falls within the range of
1.67-2.25V).
ERR 006
DC input voltage
(DC voltage is
abnormal)
High
The monitored voltage is
smaller than 2.15V or
greater than 3.3V.
1. Check the power board and related
connection cable.
2. Check the main board.
ERR 007
AC input voltage
(AC power supply
High
The monitored voltage is
smaller than 2.12V or
1. Check the power board and related
connection cable.