Alarms and Troubleshooting
45
Alarm Message
Alarm Level
Alarm Cause
Troubleshooting
Patient Side OCCL
High
1. The syringe line is
occluded.
2. A low occlusion level
is set for high-viscosity
solution infusion.
3. The system will auto
reduce the volume when
an occlusion occurs.
Tap
Stop
to eliminate the
alarm.
Identify
and
remove
the
occlusion
cause and resume the
infusion.
Standby End
Middle
The standby time is out.
Tap
Cancel
to exit the
standby state.
Reminder Alarm
Low
No key operation is
performed in 2 min after
the syringe is installed.
Tap any key to eliminate
the alarm.
Relay Failed
High
The communication is
interrupted in the relay
process or any external
interference causes relay
sequence failure.
Tap
Stop
to eliminate the
alarm. Enable the local
relay again.
9.3
Infusion Pump Faults and Troubleshooting
The following table lists the alarm message dedicated for the infusion pump and the alarm level, alarm cause,
and troubleshooting.
Alarm Message
Alarm Level
Alarm Cause
Troubleshooting
Fluid Side OCCL
High
Upstream occlusion
Tap
Stop
to eliminate the
alarm. Identify and
remove the occlusion
cause and resume the
infusion.
Air Bubble In Line
High
1. There is air bubble
inside the infusion tube.
2. Damaged infusion set
is installed on the air
bubble sensor.
Tap
Stop
to eliminate the
alarm. Check that the part
of
the
infusion
set
installed on the air bubble
sensor does not deform,
and
remove
the
air
bubble.
Содержание HP-60
Страница 30: ...Maintenance 30 Please follow local laws to dispose the old battery...
Страница 35: ...Calibration 35 Schematic diagram...
Страница 88: ...Appendix 88 6 Electronic scale 500g 0 01g...