Calibration
37
8.5
Pump Head Pressure Adjustment
1
Connect the infusion tube directly to the pressure gage (10 Bar sensor), and set the unit to Bar.
2
Install the infusion tube onto the infusion pump.
3
Choose
HOME
>
Local Set
>
Maintenance
, input the password
1666
, and select the calibration mode.
4
Set the infusion rate of the infusion pump to 100 mL/h, perform purge, and start infusion.
5
Use a slot type screwdriver to adjust the position of the peristalsis pump (anticlockwise rotation increases
the pressure) until the maximum reading on the pressure gage becomes steady and falls within the range of
2.4-3.0 Bar.