MODUS
®
SURGIC AL TECHNIQUE – STEP BY STEP
2
Intermaxillary Fixation System IMF
Страница 1: ...MODUS SURGICAL TECHNIQUE STEP BY STEP MODUS 2 Intermaxillary Fixation System IMF...
Страница 2: ...Implants 9 Surgical Technique 9 IMF Screws Application 9 Determining the Screw Position 10 Inserting the Screws 11 Intermaxillary Fixation with Ligature Wires 11 Intermaxillary Fixation with Elastomer...
Страница 3: ...pected infection at or near the implantation site Known allergies and or hypersensitivity to implant materials Inferior or insufficient bone quality to securely anchor the implant Patients who are inc...
Страница 4: ...com System Overview MODUS 2 IMF is available in the following designs Description Examples 2 0 IMF SpeedTip Screws without Plateau M2 5248 08 M2 5248 11 M2 5248 14 2 0 IMF SpeedTip Screws with Platea...
Страница 5: ...hoice of the suitable implant for the specific case Description Examples Perioperative fixation of the occlusion during orthognathic or orthodontic surgery with IMF screws Perioperative fixation of th...
Страница 6: ...in and extract it vertically from the container Cutting the Titanium Arch Bar A conventional wire cutter e g Aesculap DP560R can be used to cut the arch bar Visually check the desired cutting line bef...
Страница 7: ...r Code 2 0 Blue Drills for SpeedTip IMF screws 2 0 drill 1 5 Dental Stryker M2 3159 M2 3169 25 mm Caution It is recommended not to exceed the maximum speed of 1 000 revolutions per minute Higher speed...
Страница 8: ...anium arch bar conventional wire cutters and forceps can be used Notice Only original MODUS 2 instruments are recommended for the explantation of MODUS 2 IMF screws Notice The screw will not hold with...
Страница 9: ...Position Screw placement sites are selected considering anatomical structures i e root apices neurovascular bundles or nasal mucosa and fracture positions In the maxilla the screws are placed above or...
Страница 10: ...hich in most cases makes pre drilling unnecessary Insert the first screw in the desired position Insert the remaining screws Notice An unexpected increase of the tightening torque in cortical bone can...
Страница 11: ...axillary Fixation with Ligature Wires Thread the ligature wire maximum diameter 0 7 mm through the through bore of the screw For easy orientation the through bores are placed parallel to the markings...
Страница 12: ...rmation on the application of the MODUS 2 IMF titanium arch bar MODUS 2 IMF Titanium Arch Bar Application M 4450S Titanium Arch Bar 129 mm 2 Pkg Shorten the arch see section Cutting the Titanium Arch...
Страница 13: ...System IMF 13 www medartis com Fix the arch bar with three ligature wires per quadrant Shorten the ligature wires Notice For shortening of the ligature wires any wire cutter can be used Tighten and b...
Страница 14: ...dibular teeth The next steps are analogical to the adaptation of the arch bar to the maxilla Intermaxillary fixation with elastomeric ligatures or ligature wires Notice If individual pins have contact...
Страница 15: ...refer to the MODUS 2 IMF Product Information also available at www medartis com Screws Art No M2 5248 08 M2 5248 08S M2 5248 11 M2 5248 11S M2 5248 14 M2 5248 14S M2 5249 08 M2 5249 08S M2 5249 11 M2...
Страница 16: ...s com Disclaimer This information is intended to demonstrate the Medartis portfolio of medical devices A surgeon must always rely on her or his own professional clinical judgement when deciding whethe...