Introduction
Basic requirements
GA 5752 2135 GB 18
13
1
Technical specifications
• Shore hardness of 60
• Inner diameter 6–8 mm
• Length 1.3–3.0 m
•
Vacuum resistant down to −95 kPa (must not collapse)
Prerequisites:
• The outer diameter of the tube connector on the patient side of the septic fluid jar must match
the inner diameter of the suction tube.
1.6.5.5
Utensil
Suction catheters or lances, for example, are referred to as utensils. The utensils are used to
extract septic fluids.
Prerequisites
• The inner diameter of the utensil's connector must match the outer diameter of the suction
tube.
• The utensil must be sterilisable or a sterile single-use item.
• Biocompatibility
1.6.5.6
Mechanical overflow protection
The mechanical overflow protection device protects the product against the ingress of fluids. The
tube connector must match the vacuum connection tube.