Introduction
Basic requirements
12
GA 5752 2135 GB 18
1
1.6.5
Interface description
All devices and accessories which are combined with the tapping unit must be listed in the
accessories list or meet the specifications of the interface description. Configuration of the overall
system as well as functional testing are subject to the overall responsibility of the medical staff.
Functionality and suitability of the connected accessory for each intended application must be
checked by the operator before every use. This includes the functionality of the connector
components, airtightness and suitability regarding material properties, working pressure and flow
rate.
1.6.5.1
Vacuum connection tube
The vacuum connection tube is used to connect the tapping unit and the septic fluid jar.
Technical specifications
• Shore hardness of 60
• Inner diameter 6 mm
• Length 50 cm (± 10 cm)
•
Vacuum resistant down to −95 kPa (must not collapse)
Prerequisites
• The inner diameter of the vacuum connection tube should match the outer diameter of the
tube connector on the septic fluid jar cap of the pump.
The vacuum connection tube will be referred to only as ‘connection tube’ below.
1.6.5.2
Hydrophobic bacterial and viral filter
In its function as overflow protection device, the hydrophobic bacterial and viral filter protects the
product against the ingress of particles, fluid and foam. In its function as bacterial and viral filter, it
protects the product against the ingress of bacteria and viruses.
Prerequisites
•
Pore size ≤ 1.0 μm
• The tube connector must match the tube being used.
• The hydrophobic bacterial and viral filter must close tightly against water passage at an abso-
lute pressure of up to 10 kPa.
• If required, observe the direction of flow (see note on the hydrophobic bacterial and viral filter).
1.6.5.3
Septic fluid jar including septic fluid jar cap
The septic fluid jar and septic fluid jar cap are used to collect the secretions extracted.
Prerequisites
• Low leakage.
• Always fasten the septic fluid jar securely.
• The outer diameter of the tube connector on the patient side should match the inner diameter
of the suction tube.
1.6.5.4
Suction tube
The suction tube acts as the connection between the septic fluid jar and the utensil.
Содержание FINA VAC D 150