8.6 Drying and Lubrication
8.6.1 Required Materials
• Compressed air;
•
Soft, low-lint cloth;
• FDA approved medical grade lubricant.
Thoroughly dry all parts of the scaler
handpiece, scaler nose cone, and light
pipe by blowing them with compressed
air;
1
CAUTION
: The scaler handpiece
electrical contacts must be dry at the
start and end of the sterilisation cycle,
before the cord is connected to the
device. Always make sure that the
electrical contacts of the connector
are perfectly dry, if necessary dry them
blowing with compressed air.
CAUTION
: Before starting the
sterilisation cycle, make sure that
the insert tip is thoroughly dry both
internally and externally. To do this,
blow compressed air both externally and
through the internal passage hole. This
will prevent the appearance of stains,
streaks on the surface or oxidation inside
the insert tip.