34
multipiezo
8.3 Cleaning the Tank and Cap
The following procedure must be carried out on the tank and cap of the device.
8.3.1 Preparation
1.
Disconnect the tank from the machine
);
2.
unscrew the cap from the tank.
CAUTION
: Do not sterilise the tank and
cap in an autoclave. They could be damaged.
8.3.2 Required Materials
•
Water;
•
Cleaning solution (pH 6-9);
•
Clean, soft, low-lint cloth;
• Demineralised water.
CAUTION
: To disinfect the device and/
or its accessories, it is recommended the use
of water-based disinfectant solutions with
aneutral pH (pH7).
DO NOT USE
as disinfectants:
•
Alcohol-based disinfectant solutions;
•
Peracetic acid, Formaldehyde,
• Gluteraldehyde based products or other
equivalent solutions/products;
•
very alkaline products (pH > 9);
•
Products containing sodium
hypochlorite;
•
Products containing hydrogen peroxide;
•
Products containing abrasive substances;
•
very acid products (PH < 4);
•
Products containing aldehyde, amine
and/or phenols;
• Acetone;
•
Methylethylketone;
since they may discolour and/or damage the
plastic materials.
The manufacturer disclaims all liability for any
damage caused by the substances mentioned
above. In case of damage caused by those
substances, the Warranty will be void.
8.3.3 Cleaning Method
Thoroughly rinse the inside and outside
of both the tank and cap under running
water;
1