![mectron compact piezo P2K Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/mectron/compact-piezo-p2k/compact-piezo-p2k_use-and-maintenance-manual_1759164034.webp)
28
COMPACT PIEZO P2K
6.3
Drying and Lubrication
6.3.1
Required Materials
• Compressed air;
• Soft, low-lint cloth;
• Medical grade lubricant.
Thoroughly dry all parts of the scaler
handpiece and scaler nose cone, in
particular the electrical contacts by
blowing compressed air;
1
CAUTION
: The scaler handpiece
electrical contacts must be dry at the
start and end of the sterilisation cycle,
before the cord is connected to the
appliance. Always make sure that the
electrical contacts of the connector
are perfectly dry, if necessary dry them
blowing with compressed air.
CAUTION
: Before starting the
sterilisation cycle, make sure that the
insert is thoroughly dry both internally
and externally. To do this, blow
compressed air both externally and
through the internal passage hole. This
will prevent the appearance of stains,
streaks on the surface or oxidation inside
the insert.
2