12
COMPACT PIEZO P2K
4
USE
4.1
Controls
For the description of the controls and warnings, see the manual supplied by the manufacturer
of the dental unit.
4.2
Power On/Off
To turn COMPACT PIeZO P2K on and off, see the manual supplied by the manufacturer of the
dental unit.
4.3
Safety Requirements Before and During Use
WARNING
: Before starting work always
ensure that you have all replacement parts
(handpiece, inserts, keys) to use should there
be downtime due to a fault or problem.
WARNING
:
Exclusively use original
Mectron inserts, accessories and
replacement parts.
WARNING
:
Use of non-original
inserts not produced by MECTRON
: this
entails permanent damage to the thread
of the handpiece and compromises correct
operation with the risk of causing harm to the
patient.
CAUTION
:
Contraindications -
Ultrasonic Scaler.
Do not perform treatments
on metal or ceramic prosthetics. Ultrasonic
vibrations could lead to decementation of the
dentures.
WARNING
:
Contraindications.
Do not
use COMPACT PIEZO P2K on patients with
Pacemakers or other implantable electronic
devices. This regulation also stands for the
operator.
WARNING
: Do not perform scaling
treatments without water spray in order
to avoid an overheating of the insert that
can cause damage to the tooth. Treatments
without water spray can only be those
performed with the "Dry Work" inserts
without water passage.
CAUTION
: In treatments requiring
irrigation, use only inserts with a liquid
passage.
WARNING
:
Treatments that require
irrigation.
Always check the operation of the
irrigation before and during use. Make sure
that the liquid exits from the insert.
Do not use the appliance if the irrigation
does not work.
WARNING
:
Infection control. First
use:
All parts and reusable accessories (new
or returned from an Authorised MECTRON
Service Centre) are delivered in NON-STERILE
conditions and must be treated before each
use, according to the instructions provided in
Subsequent uses:
After each treatment, clean and sterilise all
parts and reusable accessories according to
CAUTION
: For correct use of the
appliance, the pedal must be pressed and
the appliance started with the insert not in
contact with the part to be treated, so that
the electronic circuit is able to recognise
the best point of resonance of the insert
without interference, thus allowing optimum
performance.
WARNING
: Before each treatment
ensure that the insert that is suitable for the
treatment is inserted on the handpiece.
Use only the Mectron torque wrench to
secure the insert to the handpiece.
WARNING
: Do not change the insert
whilst the handpiece is in operation in order
to avoid causing injury to the operator.