DIGITAL CLAMP METER

MODEL : 

72-AUTO/72T-AUTO

INSTRUCTION  MANUAL

®

1.  SPECIFICATIONS

1.1 General Specification
Display

n

3¾ digit liquid crystal display (LCD) with a
maximum reading of 4000 counts.

Polarity

n

Automatic, positive implied, negative polarity
indication.

Overrange

n

(OL) or  (-OL)  is displayed.

Zero

n

Automatic.

Low battery

n

“       ” is displayed when the battery

indication

voltage drops below the operating level.

Measurement

n

3 times per second, nominal.

rate
Operating

n

0°C to 50°C at < 70% relative humidity.

Environment
Storage

n

-20°C to 60°C, 0 to 80% R.H. with battery

Temperature

removed from meter.

Accuracy

n

Stated accuracy at 27

0

C ± 5

0

C,

<75% relative humidity.

Power

n

Two  1.5V  ‘AAA’  Size Battery

Battery life

n

200 hours typical

Dimensions

n

185 x 65 x 28mm (approx.)

Weight

n

170gms  including battery (approx.)

Accessories

n

Pair of test leads x 1, Instruction manual x 1,
1.5V batterys (installed) x 2,
Carring Case x 1, (K type thermocouple upto
260

0

C  for 72T-AUTO only) x 1

Maximum Jaw

n

25mm

Opening

1.2 Electrical Specification

Accuracies are ± (% r number of digits) at 27 ± 5°C and
humidity of less than 75% RH.

AC CURRENT (Auto Ranging) 50-60Hz

DC VOLTAGE (Auto Ranging)

Range

Accuracy

Overload Protection

600V

±(0.8% rdg + 3 dgt)

600V DC / AC rms

Range

Accuracy

Overload Protection

600V

±(1.2% rdg + 3 dgt)

600V DC / AC rms

RESISTANCE (Auto Ranging)

Range

Accuracy

Overload Protection

40M

V

V

V

V

V

±(1.2% rdg + 3 dgt)

250V DC / AC rms

AC VOLTAGE (Auto Ranging) 50-60Hz

Frequency (Auto Ranging)

Range : 10Hz, 100Hz, 1000Hz, 10KHz, 100KHz, 1000KHz, 10MHz
Accuracy : ± (0.5% rdg + 2 dgt)

Sensitivity : Approx 0.7V

Overload Protection : 250V DC / AC rms

Continuity Check

Threshold Level

: 40

V

 Approx.

Response Time

: 1m Sec. Approx.

Open Circuit Voltage : 0.4V Approx.

Indication

: ‘   ’ is displayed on LCD and buzzer sounds

at continuity.

3

2

1

Range

Accuracy

Overload Protection

40A

±(2.5% rdg + 5 dgt)

400A AC Max.

400A

for 1 minute

Diode Test

Measurement Current : 1.0 ± 0.6 mA Approx.
Open Circuit Voltage

: 1.6V Approx.

Capacitance (Auto Ranging) (72T-AUTO only)

Range

: 5nF, 50nF, 500nF, 5

m

F, 50

m

F, 200

m

F.

Accuracy

: ± (3% rdg + 2 dgts)

Overload Protection : 250V DC / AC rms

Temperature (72T-AUTO only)

Range

: -20

0

C to 750

0

C

Accuracy

: ± (3% rdg + 5 dgts)

Resolution

: 1

0

C

Sensor

: K type thermocouple

2. OPERATION

Before taking any measurements, read the safety information section.
Always examine  the instrument for damage, contamination (excessive
dirt,  grease, etc.) and defects. Examine the test leads for cracked or
frayed insulation. If any abnormal conditions exist do not attempt to
make any measurements.

AUTO POWER OFF (APO)

The meter will be switch off if no range switch or key is used for approx
15 minutes.

FUNCTION BUTTON

The Button is used for select 

V

 / 

 /   /   (72T-AUTO only) range.

HOLD BUTTON

Press Hold button to toggle in and out of Hold mode, In the Hold mode,
the “ H ” annunciator is displayed.

REL BUTTON

Press “REL” button to toggle in and out of “REL” mode. In this mode