13.
Motor Replacement Parvalux
39. Remove the base plate as per 7, Instructions 2 to 6
40. The motor/gearbox unit, complete
with an encapsulated encoder
(also known
as the ‘tachometer’),
is attached to the base casing via
a mounting plate retained with 3-
off M6 x 12 cap head screws.
41. Locate the motor as identified in
the photograph
42. Remove the Transformer as per 8,
43. Remove the motor red/black
connections to the PCB:
Red from J10, Black from J11.
44. Record the connections before
removing - refitting the wrong way
round will cause stand to go in