M-142 Multifunction Calibrator
MEATEST, s.r.o.
44
User Manual v54
Simultaneous functions
The multifunction calibrator allows simultaneous generation of calibrated signal along with measurement of
another signal using the built-in multimeter. To use both parts of the calibrator simultaneously, cable adapters
supplied by the manufacturer are necessary.
When using the calibrator, the following states are possible; each state has its limitation, listed in the table.
Cable adapter applied
Limitation of output signals
Limitation of multimeter’s
function
Method of connection
1
No cable adapter is used.
Calibrator output signals are not
limited in any way and can be fully
used.
Calibrator output signals are
available at the terminals located at
the front panel.
The multimeter cannot be
used.
If the multimeter is activated,
FAIL error message appears.
2
The calibrator is used with
140-01 cable adapter.
Calibrator output signals are not
limited in any way and can be fully
used.
Calibrator output signals are
available only at the terminals
located at the cable adapter.
The multimeter can only be
used to measure the external
temperature using a Pt100
sensor built into the adapter.
140-01 cable adapter
connected onto calibrator’s
terminals.
3
The calibrator is used with
Option 40 Canon – 2 banana
cable end.
Calibrator output signals are not
limited in any way and can be fully
used.
Calibrator output signals are
available on the terminals located
at the front panel.
The multimeter can be used in
the following ranges:
- DC voltage up to 12 V
- DC voltage up to 25 mA
- frequency up to 15 kHz.
Option 40 Cable, connected
onto calibrator’s
AUXILIARY connector.
4
The calibrator is used with
Option 60 Canon – 4 banana
cable end..
Calibrator output signals are not
limited in any way and can be fully
used.
Calibrator output signals are only
available at the terminals located at
the front panel.
The multimeter can only be
used in the following ranges:
- temperature using TC
sensors
- temperature using RTD
sensors
- resistance up to 2 kOhm
Option 60 cable end ,
connected onto calibrator’s
AUXILIARY connector
5
The calibrator is used with
Option 70 Canon – 4
terminals.
Following output signals can be set
only:
- resistance in four-wire
connection
– resistance temperature sensor
simulation in four-wire connection
Output signals are available on
adapter Option 70 only.
Multimeter cannot be used.
If the multimeter is activated,
FAIL error message appears.
Cable adapter Option 70,
installed on AUXILIARY
connector.
6
The calibrator is used with
Option 80 Canon – 2 terminals
Calibrator output signals are not
limited in any way and can be fully
used.
Calibrator output signals are
available on the terminals located
at the front panel.
The multimeter can be used in
the following ranges:
- mVDC voltage up to 2 V
- TC temperature
measurement using external
TC sensor
Option 80 Cable, connected
onto calibrator’s
AUXILIARY connector.
7
The calibrator is used with
Option 90 Pt100 external
temperature sensor.
Calibrator output signals are not
limited in any way and can be fully
used.
Calibrator output signals are
available on the terminals located
at the front panel.
The multimeter can be used in
the following ranges:
- RTD temperature
measurement only
Option 90 Cable, connected
onto calibrator’s
AUXILIARY connector.
Tab. 5 Cable adapters description