M-142 Multifunction Calibrator
MEATEST, s.r.o.
40
User Manual v54
•
Press the knob of the potentiometer to complete the setting.
•
After the unit of measurement is selected, use
or
cursor buttons to move onto the next menu option. The
same can be achieved using the potentiometer after pressing the knob. Press EXIT display button the close
the range selection menu and to return to the basic display.
To reset the unit of measurement to original setting, press CLR display button. Original settings are listed below:
*
basic DC voltage range VDC
V
*
DC current mA DC 20 mA
mA
*
small DC voltages mV DC up to 100 mV
mV
*
four-wire resistance measurement
*
frequency up to 15 kHz
Hz
*
temperature measurement using
thermocouples
°C
*
temperature measurement using resistive
temperature sensors
°C
Use of calculation formula
Each function of the multimeter can display recalculated measured value. The value shown on the display is
always calculated according to the formula:
Y = A0 + A1*X + A2 * X
2
where
X is the value measured by the multimeter
Y is the value shown on the display
Default setting of the coefficients is A0 = 0, A1 = 1, A2 = 0. This means that directly measured value is
displayed on the display. If any other recalculation is required, enter the desired parameters. Parameter A0 must
be of the same dimension as is in row UNIT.
•
Press METER direct control button and select A0 (A1, A2) option from the menu using cursor buttons.
•
Enter new value using numeric keyboard.
•
Confirm the value by pressing ENTER.