![Meatest 9010+ Скачать руководство пользователя страница 62](http://html.mh-extra.com/html/meatest/9010/9010_user-manual_1758451062.webp)
DCV range summary:
0.00000 mV – 1050.000 V
ACV range summary:
1.00000 mV
rms
– 1050.000 V
rms
Voltage ranges:
auto, 20 mV, 200 mV, 2 V, 20 V, 100 V, 280 V, 1050 V
Range limits:
150 – 1050 V for 1050V range, 5 – 100 % of range otherwise
Available AC units:
RMS, peak, peak-peak, average
Frequency range:
15.000 Hz – 300.00 kHz
below 2 V
15.000 Hz – 100.000 kHz
for 2 V – 20 V
15.000 Hz – 30.000 kHz
for 20 V – 200 V
15.000 Hz – 10.0000 kHz
for 200 V – 280 V
20.000 Hz – 1000.00 Hz
above 280 V
Frequency uncertainty:
10 ppm
Voltage output modes:
passive 50Ω output, available in DC up to 200 mV
active output, available in entire range of both DC and AC
Содержание 9010+
Страница 1: ......
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ......
Страница 12: ...See 9010 SCPI manual for complete SCPI reference more details on communication setup and troubleshooting ...
Страница 33: ...Following items can be modified ...
Страница 46: ......
Страница 47: ......
Страница 48: ......
Страница 49: ......
Страница 50: ......
Страница 51: ......
Страница 55: ......
Страница 57: ......
Страница 58: ......
Страница 78: ...Temperature range 250 00 1820 00 C Cold junction comp modes Manual ...