In Harmonic mode any distortion represented with amplitude and phase shift can be set in output signal.
Harmonic content of output voltage can be set independently.
Select Power mode using POWER key. Push Harmonic soft key. Following screen will appear:
See chapter 4.1.3 for individual harmonic component editing, previewing, saving, and loading. Star sign
in field Harmonic indicates that for the output signal has been defined harmonic distortion. Label without
star sign means that the output signal is pure sin wave.
In Energy mode calibrator is sourcing output signal for defined period. Select Power mode using
POWER key. Push Energy soft key. Following screen will appear:
Set requested time for which output signals should be connected to the output terminals. Use SELECT
soft key to move active window. Both SINE and Harmonic mode can be applied in Energy mode.