Only output voltage and current can be set in DC mode. In AC mode frequency of the output signal and
phase shift between voltage and current can set as well.
1.
Connect power meter to the calibrator as shown on the figure below.
2.
Push POWER function key on the calibrator. Push POWER key once more to switch between
DC and AC mode.
3.
Set output voltage, current, frequency and phase shift using soft keys beneath the display or
switch active field on the display for direct value entry using SELECT keys.
4.
Push OPER key to connect output terminals. Calibrator is sourcing voltage and current.
5.
Push STBY key to switch output off.
Calibrator features with floating current channel. Setting of grounded or floating output can be done in
SETUP menu.
Sign of AC wave indicates SIN mode without distortion.
Terminals
HI
LO
I+
I-
HIR
GND
Entry
Function
TRG
HF
9010
Multifunction Calibrator
V
COM
A
R
METER
SCOPE
SOURCE
METER
NULL
SELECT
SETUP
SYSTEM
1100V
RMS
MAX
1500V
DC
MAX
1V PK
MAX
20V PK
MAX
1 00V PK
MAX
10V PK
MAX
2V PK
MAX
25V PK
MAX
10V PK
MAX
50V PK MAX
50 V PK MAX
FUNC
OPER
STBY
10V
RMS
MAX
Содержание 9010+
Страница 1: ......
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ......
Страница 12: ...See 9010 SCPI manual for complete SCPI reference more details on communication setup and troubleshooting ...
Страница 33: ...Following items can be modified ...
Страница 46: ......
Страница 47: ......
Страница 48: ......
Страница 49: ......
Страница 50: ......
Страница 51: ......
Страница 55: ......
Страница 57: ......
Страница 58: ......
Страница 78: ...Temperature range 250 00 1820 00 C Cold junction comp modes Manual ...