Chapter 6
112
For example, if you are using 512 channels in the channel list, scanning the channel list 256
times every trigger or retrigger, and using an A/D sample clock with a frequency of 1 MHz,
set the maximum retrigger frequency to 7.62 Hz, since
7.62 Hz =
1
( 512 * 256) +2
μ
s
1 MHz
To select software-retriggered scan mode, use software to specify the following parameters:
• The dataflow as Continuous, ContinuousPreTrigger, or ContinuousPrePostTrigger,
• Triggered scan mode usage as enabled.
• The retrigger source as Software.
• The number of times to scan per trigger or retrigger (also called the multiscan count).
• The frequency of the Triggered Scan Counter.
The initial trigger source depends on the trigger acquisition mode selected; refer to
for more information on the supported trigger acquisition modes and trigger sources.
Note:
An A/D Trigger Out signal is provided for your use. This signal is high when the
A/D subsystem is waiting for a trigger and low when a trigger occurs. In
software-retriggered scan mode, this signal stays low until the desired number of samples
have been acquired, then goes high until the software retrigger is generated.
Externally-Retriggered Scan Mode
Use externally-retriggered scan mode if you want to accurately control the period between
conversions of individual channels and retrigger the scan based on an external event. Like
software-retriggered scan mode, this mode allows you to acquire 262,144 samples per trigger
(256 times per trigger x 1024-location channel list).
Note:
Use externally-retriggered scan mode with continuous post-trigger acquisitions only;
for more information on post-trigger acquisitions.
When it detects an initial trigger (post-trigger source only), the board scans the channel list up
to 256 times, then waits for an external retrigger to occur. Specify any supported post-trigger
source as the initial trigger. For the retrigger, specify either an external digital (TTL) trigger.
When the retrigger occurs, the board scans the channel list the specified number of times, then
waits for another external retrigger to occur. The process repeats continuously until either the
allocated buffers are filled or you stop the operation; refer to
for more information on
buffers.
Содержание Data Translation DT3010 Series
Страница 2: ...DT3010 Series UM 16866 V User s Manual Title Page ...
Страница 5: ......
Страница 15: ...About this Manual 14 ...
Страница 16: ...15 1 Overview Features 16 Supported Software 18 Accessories 19 Getting Started Procedure 21 ...
Страница 23: ...Chapter 1 22 ...
Страница 24: ...Part 1 Getting Started ...
Страница 25: ......
Страница 37: ...Chapter 2 36 ...
Страница 49: ...Chapter 3 48 ...
Страница 91: ...Chapter 4 90 ...
Страница 102: ...Part 2 Using Your Board ...
Страница 103: ......
Страница 147: ...Chapter 6 146 ...
Страница 159: ...Chapter 7 158 ...
Страница 160: ...159 8 Calibration Calibrating the Analog Input Subsystem 161 Calibrating the Analog Output Subsystem 169 ...
Страница 176: ...175 9 Troubleshooting General Checklist 176 Technical Support 178 If Your Board Needs Factory Service 179 ...
Страница 181: ...Chapter 9 180 ...
Страница 193: ...Appendix A 192 ...
Страница 213: ...Appendix B 212 ...
Страница 231: ...Index 230 ...