Adjustment of Input Sensitivity
Dimensions
Instruction Manual
15
The input sensitivity can be adjusted by moving the jumper. The input sensitivity for
position 1 is specified in the type designation. In position 5, the amplification can be
steplessly adjusted with the Trimmer “TR“.
Position Amplification
factor
Output voltage at sensitivity
option 2 mV/V
Output voltage at sensitivity
option 3.5 mV/V
1
1
±5 volts at
±2 mV/V
±5 volts at
±3.5 mV/V
2
2
±5 volts at
±1 mV/V
±10 volts at
±2 mV/V
±5 volts at
±1.75 mV/V
±10 volts at
±3.5 mV/V
3
4
±5 volts at
±0.5 mV/V
±10 volts at
±1 mV/V
±5 volts at
±0.875 mV/V
±10 volts at
±1.75 mV/V
4
10
±5 volts at
±0.2 mV/V
±10 volts at
±0.4 mV/V
±5 volts at
±0.35 mV/V
±10 volts at
±0.7 mV/V
5
1...10
±5 volts at
±0.2...2 mV/V
±10 volts at
±0.4 ...2 mV/V
±5 volts at
±0.35...3.5 mV/V
±10 volts at
±0.7 ...3.5 mV/V