ANI Monitor V1 - Continuous analgesia monitoring system
6
MD/PRD/IN16.ANIV1 V.14 - 12 MAR 2020
- CONTACT YOUR MDOLORIS MEDICAL SYSTEMS REPRESENTATIVE.
IN OPERATING ROOMS THE ANI Monitor V1 MUST BE PLACED OUTSIDE THE
EXPLOSION HAZARD ZONE.
MAKE SURE THE MONITOR IS INSTALLED OUTSIDE THE LIQUID PROJECTIONS
HAZARD ZONE, E.G. PERFUSION BAG.
NEVER MODIFY THE MONITOR WHEN OPEN.
USE OF ANI Monitor V1 ADJACENT TO OR STACKED WITH OTHER EQUIPMENT
SHOULD BE AVOIDED BECAUSE IT COULD RESULT IN IMPROPER OPERATION.
IF SUCH USE IS NECESSARY, THIS EQUIPMENT AND THE OTHER EQUIPMENT
SHOULD BE OBSERVED TO VERIFY THAT THEY ARE OPERATING NORMALLY.
PORTABLE RF COMMUNICATIONS EQUIPMENT (INCLUDING PERIPHERALS
SUCH AS ANTENNA CABLES AND EXTERNAL ANTENNAS) SHOULD BE USED NO
CLOSER THAN 30 CM (12 INCHES) TO ANY PART OF THE ANI Monitor V1,
INCLUDING CABLES SPECIFIED BY THE MANUFACTURER. OTHERWISE,
DEGRADATION OF THE PERFORMANCE OF THIS EQUIPMENT COULD RESULT.
1.2
Caution
Read this entire manual carefully before using the monitor in a clinical setting.
Before starting, ensure there are no USB devices (USB sticks for example) connected to
the monitor.
The patient should not be able to reach the equipment directly or indirectly; avoid for instance
placing equipment on top of another equipment with a metal casing.