MD/PRD/IN16.ANIMOC9 V.02 - 12/03/2020
19 | 30
Réf
Feature
Description
1
Slot 1
Displays the level and the value of the selected parameter
2
Slot 2
Displays the value of the other parameters
3
ANIm
Averaged value of the ANI between 0 and 100
4
ANIi
Instantaneous value of the ANI between 0 and 100
5
Energy
Energy of the RR series
6
Computation
quality
Quality of the measurement. Values ranging from 0 to1
• 0 = bad quality
• 1 = good quality
Mode of operation
Once the module is connected to the patient with the sensor and to the root, the calculation
algorithm will automatically begin.
Note: The ANI technology can be used with both conscious and unconscious patients,
whenever the physician wishes to use it.
ECG acquisition
Check for good computation in the lower part of the Root screen.
If the indication shows poor measurement quality (Value equal to 0) the data displayed
cannot be consider, the ANI index is not displayed anymore, the curve freezes. In that case,
check the proper position of the sensor and their adhesion to the skin. If necessary,
reposition or change them.
WARNING: always check the quality parameter. The ANI index will not be reliable if the
signal quality is bad.