ANI Monitor V2 - Continuous analgesia monitoring system
MD/PRD/IN16.ANIV2 v.07 - 12 MAR 2020
P a g e
17 | 45
Check for good ECG signal quality in the lower part of the screen. If there is no signal, check
the connection between the monitor and the electrodes. Try at least once to unplug and plug
in the connection again to recover the ECG signal. If a signal is displayed but does not look
right, check that the electrodes are properly placed and connected to the cable.
WARNING: always check in the upper left window that the signal quality is satisfactory
(green indicator). The ANI index will not be reliable if the signal quality is not good enough
(red indicator).
If the user thinks the ECG amplitude is too low, select «
Reset ECG
» to automatically
recalibrate the ECG signal acquisition.
5
Using the ANI Monitor V2 and setting parameters
5.1
ECG Capture
The lower window shows the ECG acquired by the monitor. This ECG is filtered of all
technical and physiological artifacts (e.g. premature ventricular contractions).
ECG
WARNING
:
Always check the good quality of the ECG signal by
1)
Both with the morphology signal and the scrolling on the ECG display
2)
looking at the bottom right corner of the screen :
-
Good quality signal
- Medium quality signal
-
Poor quality signal
Case n°1 - No signal is displayed
You must try, at least once, to replace the sensor.