MLA 900 Operating Instructions V 2.0 8011381 © MBA Instruments GmbH
Handling
4.5
Measures in the event of a fault/damage
b
If the MLA 900 is faulty of damaged:
send both the display unit and the measuring
probe to the manufacturer's factory for repair (detailed information
).
WARNUNG: Risks caused by damage
If a fault or damage to an instrument component is detected, the MLA 900 must
not be taken into a hazardous location.
If the MLA 900 is faulty or damaged:
b
Mark the instrument clearly as faulty (for example, using a sticker).
b
Make sure that the MLA 900 is no longer used in hazardous locations.
WARNUNG: Risk caused by damaged measuring probe
If the cylinder of the measuring probe is deformed (for example, partly dented),
the conductivity measuring values will be distorted. This fault may not be detected
during a performance test (
b
Check the mechanical condition of the measuring probe.