1
U S E R / O W N E R ’S M A N U A L
Version 1 / Revision 1
Specialists In Pressure Care™
Страница 1: ...1 U S E R O W N E R S M A N UA L Version 1 Revision 1 Specialists In Pressure Care...
Страница 2: ...ety Specifications 4 Operating Principles Installation 5 Mattress Installation Steps 3 4 5 6 Mattress Installation Step 6 7 Troubleshooting Guide 8 Cleaning Disinfection 9 Mattress Warranty 10 Warning...
Страница 3: ...ll Mayflex preventative aids must be used as part of an individualised care plan which include proper nursing practices i e Patient turning re positioning and regular patient skin assessments Professi...
Страница 4: ...ith 93 42 EEC Unique Unit Example number SNXXXXXX on pump Directive 73 23 EEC Test standard EN60335 1 A141995 Safety Standards CE Due to continual development of the above product specifications are s...
Страница 5: ...yflex logo is at the top end of the bed Fasten the Mayflex overlay securing straps so they are underneath the underlying foam mattress CAUTION Please ensure that mattress height does not render any be...
Страница 6: ...s of the mattress The pump will then begin to inflate the air cells and run continuously in dynamic mode The alternating function will inflate and deflate the mattress cells over a 10 minute cycle CAU...
Страница 7: ...deflation simply disconnect the two air hoses from the Mayflex pump unit and connect them together This will result in the mattress system being equalised and the mattress cells will then retain press...
Страница 8: ...air connections are fitted securely Ensure air is coming from pump Ensure connectors are fastened securely and reconnect pump air hoses if loose See Step 2 Pg3 Check CPR is connected correctly and is...
Страница 9: ...In many cases it will be only be necessary to remove the mattress cover for cleaning WARNING Protective clothing should be warn when performing cleaning procedures WARNING Do not use Phenol based clea...
Страница 10: ...will be sent out a loan system next working day if notified before 12pm while your mattress is repaired or replaced free of charge For extended warranty packages please contact Mayflower Medical on 01...
Страница 11: ...effectiveness of the pressure care systems provided Internal and External research is and will always be ongoing These risk factors should not be taken as prescriptive criteria WARNING Before any clea...
Страница 12: ...12 Mayflower Medical Wessex House Upper Market Street Eastleigh Hants SO50 9FD Specialists In Pressure Care Copyright Mayflower Medical 2009...