ACRYLIC FLOW METERS
A S S E M B L Y I N S T R U C T I O N S &
I N S T R U C T I O N S F O R U S E
R 2 1 9 P 8 6 ( 0 - 1 2 0 L P M )
R 2 1 9 P 8 7 ( 0 - 7 0 L P M )
R 2 1 9 P 8 8 ( 0 - 3 0 L P M )
R138P11 Rev. E
0 - 3 0 , 0 -7 0 & 0 -1 2 0 L P M
Страница 1: ...LIC FLOW METERS A S S E M B L Y I N S T R U C T I O N S I N S T R U C T I O N S F O R U S E R 2 1 9 P 8 6 0 1 2 0 L P M R 2 1 9 P 8 7 0 7 0 L P M R 2 1 9 P 8 8 0 3 0 L P M R138P11 Rev E 0 30 0 70 0 12...
Страница 2: ......
Страница 3: ...nings 1 1 4 Cautions 1 2 2 0 PRODUCT SPECIFICATIONS 3 2 1 Flow Ranges Graduations Accuracy 3 2 2 Transport Storage Requirements 3 3 0 OPERATING INSTRUCTIONS 4 3 1 General Operation 4 4 0 TROUBLESHOOTI...
Страница 4: ......
Страница 5: ...fore administering to patient and monitor flow on a frequent basis Flow meters may contain magnetic ferrous material that may affect the results of an MRI Read this User Manual before installing or op...
Страница 6: ...E if damaged Inspect the flow meter for visual damage before use DO NOT over tighten knob when turning off This will cause damage to the flow meter DO NOT over tighten any threaded connectors body may...
Страница 7: ...e output of the flow meter when the flow indicator is beyond the highest calibrated graduation The Flush Flow range is as indicated on the flow meter labeling 2 2 Transport Storage Requirements 40 F 4...
Страница 8: ...the float is not resting at the bottom of the flow tube the product is leaking consult the TROUBLESHOOTING Section 4 0 4 Adjust flow To increase Turn knob counterclockwise To decrease Turn knob clock...
Страница 9: ...Replace valve Sticking float ball Debris in flow tube Clean flow tube float ball Unable to set desired flow Blocked inlet Replace filter Knob will not turn Valve seized Replace valve 5 0 RECOMMENDED M...
Страница 10: ...ited to making replacements repairs or issuing credit for equipment found to be defective This warranty extends only to the buyer purchasing the equipment directly from Maxtec or through Maxtec s desi...
Страница 11: ...al Rubber Latex Storage Temperature Non sterile Authorized Representative non sterile Manufacturer PVC Contains no Polyvinyl Chloride Warning Maxtec 2305 South 1070 West Salt Lake City Utah 84119 USA...