Maxtec MAXO2 OM-25ME Скачать руководство пользователя страница 1

OXYGEN

MONITOR

MODEL

OM-25ME

(OM-25ME-L)

USER’S GUIDE

AND

OPERATING

INSTRUCTIONS

manufactured by:

R213M05   Rev. H

®

®

R213M05H.qxd  6/26/2006  9:04 AM  Page 1

Содержание MAXO2 OM-25ME

Страница 1: ...OXYGEN MONITOR MODEL OM 25ME OM 25ME L USER S GUIDE AND OPERATING INSTRUCTIONS manufactured by R213M05 Rev H...

Страница 2: ......

Страница 3: ...i Manufacturer Maxtec Inc European Representative Airox S A Parcd activities Pau Pyrenees L echangeur BP833 64008 PAU Cedex France...

Страница 4: ...and operation will provide you with years of exceptional service We also encourage your comments or suggestions as to how our equipment in any way can better serve your needs Please feel free to write...

Страница 5: ...o so may seriously impair the monitor s performance Repair or alteration of the MAXO2 beyond the scope of the maintenance instructions or by anyone other than an authorized Maxtec service person could...

Страница 6: ...ification Protection against electric shock Internally powered equipment Protection against water Ordinary equipment Mode of Operation Continuous Sterilization See section 6 Safety of application in t...

Страница 7: ...on 10 3 OPERATING INSTRUCTIONS 12 3 1 Alarm Setting Procedure 12 3 1 1 Low Alarm Setting 12 3 1 2 High Alarm Setting 12 3 2 Basic Operation 13 3 3 Alarm Conditions 13 4 SENSOR REMOVAL AND REPLACEMENT...

Страница 8: ...erter fitting for standard 15 mm T adapter Operation using 2 AA alkaline batteries 2 x 1 5 volts for approximately 3000 hours of performance in typical usage not in alarm conditions For extra extended...

Страница 9: ...e the high and low alarm settings Low Alarm Indicator The low alarm setting is displayed at all times just below the LOW icon on the LCD readout The indicated value represents the oxygen percentage at...

Страница 10: ...LCD Readout Alarm LEDs Keypad External Probe A C B D 1 2 3 10 12 4 13 15 16 14 6 17 18 8 OM 25ME 11 9 5 7 3...

Страница 11: ...conds the LOCKED icon will reappear and the unit will revert to the LOCKED mode When the unit is in the unlocked state pressing the LOCK UNLOCK key will also cause the LOCKED icon to reappear CALIBRAT...

Страница 12: ...e if 10 seconds elapse and no keys are pressed The OM 25ME alarm setting is adjustable down to 18 The OM 25ME L low alarm setting is adjustable to 15 HIGH SET Key When the keypad is unlocked pressing...

Страница 13: ...the teflon membrane and immediately reacts electrochemically at a gold cathode Concurrently oxidation occurs electrochemically at a lead anode generating an electrical current and providing a voltage...

Страница 14: ...dout are turned on for approximately 2 seconds On OM 25ME units the alarm beeper sounds and the high and low alarm LEDs are illuminated When the diagnostic test is completed successfully the word CAL...

Страница 15: ...nvironment in which the MAXO2 will be used A known value of oxygen is defined as an oxygen source which has a traceable certificate and or USP certification Note Before beginning calibration the MAX 2...

Страница 16: ...entered value will not allow operation within the specified output range of the sensor This situation may occur if a the operator has inadvertently entered the wrong concentration for the calibration...

Страница 17: ...erature pressure and humidity Effects of Temperature The MAXO2 Monitor will hold calibration and read correctly within 3 when in thermal equilibrium within the operating temperature range The device m...

Страница 18: ...librating as when measuring 3 The MAXO2 Monitor oxygen sensor has been validated at pressures up to 2 atmospheres absolute Calibration or operation above this pressure is beyond the intended use Humid...

Страница 19: ...larm value cannot be set lower than 18 nor can it be set closer than 1 from the high alarm value due to ASTM specifications special applications requiring alarms below 18 can be accomodated by the mod...

Страница 20: ...sor is mounted in the adapter with the flow diverter pointing downward This will prevent moisture from draining into the sensor membrane Note It is important that a tight fit exists between the probe...

Страница 21: ...flow diverter from the sensor and discard the expired sensor Note The sensor contains lead and lead acetate be sure to dispose of expired sensors in accordance with hospital local state and federal r...

Страница 22: ...t If at any time ErX i e Er1 Er4 etc appears on the LCD readout contact Maxtec s Customer Service Department 6 CLEANING AND MAINTENANCE When cleaning or disinfecting the MAXO2 Monitor take appropriate...

Страница 23: ...e Maxtec MAX 250E galvanic fuel cell Expected Sensor Life 900 000 O2 Hours over 2 years in typical applications Alarm System high low alarms flashing red LEDs nominal 2300Hz audible alarm beeper Low A...

Страница 24: ...gen the monitor will display a reading greater than the actual oxygen percentage This is because the monitor was originally calibrated at atmospheric pressure 0 psig then exposed to a higher pressure...

Страница 25: ...abilize before adjusting the display This may take 30 seconds or more Externally mounted sensors as found on the OM 25ME or OM 25ME L come equipped with diverter tips The tips help direct the gas in a...

Страница 26: ...to condense on the sensor Condensation on the sensor may eventually affect performance For this reason it is recommended that the sensor be mounted in a vertical position facing downward to prevent c...

Страница 27: ...Bracket R206P75 Monitor Analyzer Pole Mount Clamp R213P02 Monitor Analyzer Protective Carrying Case R213P31 Swivel Mount Stand RP16P02 T Adapter 15 mm I D RP16P10 T Oxygen Sensor Activator System R21...

Страница 28: ...ec or through Maxtec s designated distributors and agents as new equipment Maxtec warrants the MAX 250E oxygen sensor in the MAXO2 Monitor to be free from defects in material and workmanship for a per...

Страница 29: ......

Страница 30: ...Maxtec Inc 6526 South Cottonwood Street Salt Lake City UT 84107 General Tel 801 266 5300 Toll Free Dial 800 748 5355 FAX 801 270 5590 Home Page http www maxtecinc com email sales maxtecinc com...

Отзывы: