Table 4. Format of Received Samples—WSpO

2

 in 

Calibration/Algorithm Mode 

 
 

 

 

DATA SOURCE 

BYTE 

INDEX  DATA ITEM 

NUMBER OF 

BYTES 

(MSB FIRST) 

DESCRIPTION 

MAX86141 

PPG Data 

(18 Bytes) 

LED1 

N/A 

LED2 

N/A 

LED3 

N/A 

LED4 

Red LED counter 

12 

LED5 

IR LED counter 

15 

LED6 

N/A 

Accelerometer 

(6 Bytes) 

18 

accelX 

Two's complement. LSB = 0.001g 

20 

accelY 

Two's complement. LSB = 0.001g 

22 

accelZ 

Two's complement. LSB = 0.001g 

SpO

2

 Algorithm 

(9 Bytes) 

24 

10x calculated R value 

26 

Confidence 

Calculated confidence in % 

27 

SpO

2

 

10x SpO

2

 % 

29 

% complete 

Calculation progress in % (only in 

one-shot mode of algorithm) 

30 

Low SNR 

flag 

Shows low SNR in measurement 

31 

Motion flag 

Shows excessive motion 

32 

Status 

Reported status of algorithm: 

LED adjustment: 0 

Computation: 1 

Success: 2 

Timeout: 3 

Содержание MAX32664C

Страница 1: ...ed for measurement of SpO2 and heart rate Combined with the MAX86141 optical sensor and a 3 axis accelerometer it provides the sensor s raw data as well as calculated SpO2 or heart rate data to a host...

Страница 2: ...llection in Algorithm Mode 11 2 2 2 Pure Raw Data Collection 13 2 3 Algorithm Mode 15 2 3 1 Algorithm Settings and Configurations 16 3 Measuring Heart Rate on Wrist WHRM 17 3 1 Setup 17 3 2 Raw Data C...

Страница 3: ...Table 6 Host Commands WSpO2 Pure Raw Data 13 Table 7 Format of Received Samples WSpO2 in Pure Raw Mode 14 Table 8 Host Commands WSpO2 Algorithm Mode 15 Table 9 Configurations and Settings WSpO2 16 Ta...

Страница 4: ...to properly capture and calculate the data this solution requires an accelerometer The MAX32664C firmware includes the required drivers for the Kionix KX122 accelerometer which is wired together with...

Страница 5: ...o interrupt the host for I2C communication The MAX32664C interfaces to the MAX86141 optical sensor through the SPI bus An accelerometer is mandatory for heart rate monitoring A KX122 accelerometer can...

Страница 6: ...The write transaction was successful 0x01 ERR_UNAVAIL_CMD Illegal Family Byte and or Command Byte was used 0x02 ERR_UNAVAIL_FUNC This function is not implemented 0x03 ERR_DATA_FORMAT Incorrect number...

Страница 7: ...fore starting the algorithm as described later in this document Alternatively a host side accelerometer can be used In this case the host is required to use a 3 axis accelerometer at 25Hz sampling and...

Страница 8: ...d they need to be loaded to the MAX32664C every time prior to starting the algorithm But first they are required to be converted to 32 bit integer format using the following Aint32 round 105 x a Bint3...

Страница 9: ...ettings and configurations Table 9 from default should appear here BEFORE enabling algorithm 1 7 AA 52 05 01 Enable WSpO2 algorithm AB 00 READING SAMPLES Host reads samples upon receiving MFIO interru...

Страница 10: ...celerometer 6 Bytes 18 accelX 2 Two s complement LSB 0 001g 20 accelY 2 Two s complement LSB 0 001g 22 accelZ 2 Two s complement LSB 0 001g SpO2 Algorithm 9 Bytes 24 R 2 10x calculated R value 26 Conf...

Страница 11: ...other fields can be ignored AB 00 1 3 AA 10 01 05 Set sensor hub interrupt threshold AB 00 1 4 AA 44 00 01 00 Enable AFE e g MAX86141 with sensor hub samples AB 00 1 5 AA 44 04 01 00 if sensor hub ac...

Страница 12: ...ored in the FIFO nn samples 33 bytes each will be read The format of samples is shown in Table 4 AB 00 data_for_ nn_samples STOP Host ends the procedure 3 1 AA 44 00 00 Disable AFE e g MAX86141 AB 00...

Страница 13: ...ll scale Reduce 7F if signal is saturated AB 00 1 9 AA 40 00 25 7F Set MAX86141 LED3 red current to half of full scale Reduce 7F if signal is saturated AB 00 1 10 AA 40 00 2A 3C Set MAX86141 LEDs 2 an...

Страница 14: ...ITEM NUMBER OF BYTES MSB FIRST DESCRIPTION MAX86141 PPG Data 18 Bytes 0 LED1 3 N A 3 LED2 3 N A 6 LED3 3 N A 9 LED4 3 Red LED counter 12 LED5 3 IR LED counter 15 LED6 3 N A Accelerometer 6 Bytes 18 a...

Страница 15: ...AB 00 1 5 AA 44 04 01 00 if sensor hub accelerometer is used AA 44 04 01 01 if host accelerometer is used Enable accelerometer with sensor hub or host side accelerometer AB 00 1 6 AA 50 05 02 00 Set...

Страница 16: ...rised of three 32 bit signed values 0xFFE69196 000CB735 00989680 0x01 Sampling rate Hz 0x00 100Hz 0x01 25Hz 0x01 0x02 Algorithm run mode 0x00 continuous 0x01 one shot 0x02 calibration to collect data...

Страница 17: ...ct raw PPG samples There are two ways to accomplish this which are detailed as follows 3 2 1 Raw Data Collection in Algorithm Mode In Algorithm mode automatic exposure control AEC and skin contact det...

Страница 18: ...overwritten By default WHRM sets the following AFE registers Sample rate is 100Hz 4 sample averaging Integration time 117 s ADC 1 and 2 range 32 A LED 1 full range 124mA 1 9 AA 40 00 23 7F Set the MA...

Страница 19: ...6 LED3 3 N A 9 LED4 3 Green 2 counter 12 LED5 3 N A 15 LED6 3 N A Accelerometer 6 Bytes 18 accelX 2 Two s complement LSB 0 001g 20 accelY 2 Two s complement LSB 0 001g 22 accelZ 2 Two s complement LSB...

Страница 20: ...ull scale Reduce 7F if signal is saturated AB 00 1 8 AA 40 00 2A 03 Set MAX86141 LED1 current full range 124mA AB 00 1 9 AA 40 00 20 21 Set MAX86141 LED sequence LED1 green LED2 IR AB 00 1 10 AA 40 00...

Страница 21: ...EX DATA ITEM NUMBER OF BYTES MSB FIRST DESCRIPTION MAX86141 18 Bytes 0 LED1 3 Green 1 counter 3 LED2 3 N A 6 LED3 3 N A 9 LED4 3 Green 2 counter 12 LED5 3 N A 15 LED6 3 N A Accelerometer 6 Bytes 18 ac...

Страница 22: ...ommand to change the algorithm settings and configurations Table 15 from default should appear here BEFORE enabling the algorithm Example To run algorithm in AGC Only mode no AEC set motion magnitude...

Страница 23: ...RATION INDEX DESCRIPTION DEFAULT VALUE MSB FIRST 0x50 for write 0x51 for read 0x02 0x0B AEC algorithm enable 0x00 Disable 0x01 Enable 0x01 0x0C SCD enable 0x00 Disable 0x01 Enable 0x01 0x0D Adjusted t...

Страница 24: ...SUME LIABILITY FOR OR PROVIDE A REPRESENTATION OF ACCURACY OF THE INFORMATION DEVICES OR TECHNOLOGY DESCRIBED IN THIS DOCUMENT MAXIM ALSO DOES NOT ASSUME LIABILITY FOR INTELLECTUAL PROPERTY INFRINGEME...