Maxair O.N. 2602-06 Скачать руководство пользователя страница 1

P/N 01423214 Rev A   Page 1

Bio-Medical Devices Intl  

 17171 Daimler Street    Irvine, CA 92614    800-443-3842    www.maxair-systems.com

2602-06 and 2602-06B  

6-Gang Charger and 6-Gang Bracket 

Instructions for Use

              
                 

WARNING

Use only if package is received unopened and contents are  

undamaged. If damage is noted, contact the shipper for replace-

ment or repair.

 

Prior to using any MAXAIR

®

 System or component, be sure to be 

familiar with the system’s NIOSH approved configuration.

 

DO NOT use if any component is damaged.  If any components 

are damaged or contaminated and therefore unfit for safe and  

effective use, they should be replaced immediately.
Only trained and experienced personnel who have read and  

understand the User’s Instructions should use this product.

 

Failure to follow the User’s Instructions, these and in the MAXAIR 

Systems User’s Instructions, P/N 03521015, may be hazardous to 

the user’s health.

 

The institution using this product in any application is responsible 

for determining the appropriateness of this equipment relative to 

regulatory requirements. Bio-Medical Devices Intl, Inc. does not 

recommend the appropriate systems for a particular institution or 

facility.

 

Use only MAXAIR Systems/ NIOSH approved compatible  

components.

 

NOT for use in atmospheres immediately dangerous to life or 

health (IDLH), and atmospheres containing less than 19.5% oxy-

gen, or more than 25% oxygen.
If you need more information, contact your BMDI Sales  

Representative, or call BMDI customer service at 1-800-443-3842.

 CAUTION 

 

Before use, refer to the CAPR

®

 User’s Instructions, P/N 03521015 (received with all CAPR Helmets) for System Details.

 

      

               

!

Symbol Definitions

                    

                  Warning, Caution, or Note

                  Order Number                                   Part Number

 

      

               

!

O.N.

P/N

 

      

               

!

Intended Use

The MAXAIR Systems 6-Gang UL/CE Charger (2602-06) and 

Charger Bracket (2602-06B) are intended for use only when 

charging MAXAIR Systems’ Lithium-Ion Batteries.
The 6-Gang Charger includes six 2600-02 UL/CE Chargers and 

is capable of charging up to six MAXAIR Systems’ Lithium-Ion 

batteries simultaneously.
The Charger Bracket requires from one to six MAXAIR 2600-02 

Chargers to be assembled with it to allow charging of from one to 

six MAXAIR Lithium-Ion batteries.

Specifications

ELECTRICAL SPECIFICATIONS

ELECTRICAL OUTPUT

ELECTRICAL INPUT

VOLTAGE: UP TO 16.8 VDC VOLTAGE: 100-240 VAC,  

                   50-60Hz max.

CURRENT: UP TO 5.4 A

CURRENT: UP TO 3.0 A

Recommended System Temperature Limits: 

Use/Handling: 0

o

C to 54

0

C at a maximum 80% Relative Humidity.

Charging: 0

o

C to 45

0

C at a maximum 80% Relative Humidity.

Storage: 0

o

C to 35

0

C at a maximum 80% Relative Humidity.

O.N. 2602-06 

6-Gang Charger

O.N. 2602-06B 

6

-Gang  

Charger Bracket  

(for 2600-02 Chargers)

Charging Protection from Electrical Surges

It is highly recommended to always connect MAXAIR Chargers 

directly to a Surge Protection Device, adequate for all 

anticipated occurrences, during all charging activities of MAXAIR 

LIBs (Li-Ion Batteries), and whenever the Charger is connected to a 

mains power source.
To choose an appropriate surge protector you should consult with 

your Engineering department regarding specifics to your physical 

plant and geographical environment. 

NOTE

For general safety and sensible use it is recommended          

to connect and disconnect electrical devices in a  

standard sequence. The full connecting and disconnecting  

sequence for items required for battery charging is as  

follows.

Connecting:

1. Connect the surge protector to the main (wall) power 

source and power it on.

2. Connect the Charger to the surge protector.

3. Connect the Battery to the Charger.

Disconnecting:

1. Disconnect the Battery from the Charger.

2. Disconnect the Charger from the surge protector.

3. Turn off and then disconnect the surge protector from 

the main (wall) power source.

 

      

               

!