P/N 01023649 Rev B Page 1
DLC Cuffs
(Disposable Lens Cuffs)
User Instructions
Intended Use
The MAXAIR DLC is to be donned with a MAXAIR CAPR
®
Helmet and a MAXAIR Filter Cartridge.
CAPR DLC Systems filter aerosolized and droplet particulates.
DLCs are designed for single use applications.
O.N. 2365-02SM
O.N. 2365RB-02SM*
(01031316)
Small-Medium
O.N. 2365-02ML
O.N. 2365RB-02ML*
(01031291)
Medium-Large
*RB indicates plastic
reusable dispensing box vs.
standard cardboard.
1. 2081-03 Helmet with
2051-07 SnapOn Cage
removed, 2071-08 Liner, and
2590-05 Power Cord
2. 2167-10 Filter Cartridge*
3. 2061-08 Filter Cover
Cap (FCC)
4. 2500-36TSC Battery**
5. 2000-76 Battery Belt
6. 2600-01 Battery Charger***
7. 2365-02 DLC ML & SM
Materials
DLC Lens is PETG
DLC Cuff is Polyurethane
* Alternate Filters include the 2180-05, 2166-10, and 2164-10
** Alternate Batteries include the 2500-37TSC
*** Alternate Charger includes the 2602-01.
5
6
4
3
2
7
Symbol Definitions
Warning, Caution, or Note
Order Number Part Number
!
O.N.
P/N
1
Standard CA-DLC-CAPR-36 System
1
with DLC
MAXAIR Recommended System Temperature Limits:
Use/Handling: 0
o
C to 54
0
C at a maximum 80% Relative Humidity.
Charging: 0
o
C to 45
0
C at a maximum 80% Relative Humidity.
Storage: 0
o
C to 35
0
C at a maximum 80% Relative Humidity.
All MAXAIR Systems, components, and disposables are latex free.
Regulatory
NIOSH
Specifications
WARNING
Failure to follow User Instructions P/N 03521015 and the instructions
contained herein may be hazardous to the user’s health.
Use only if package is received unopened and contents are
undamaged. If damage is noted, contact the shipper for replacement
or repair.
Prior to using any MAXAIR
®
System or component, be sure to be
familiar with the system’s NIOSH approved configuration.
DO NOT use if any component is damaged. If any components are
damaged or contaminated and therefore unfit for safe and effective
use, they should be replaced immediately.
Only trained and experienced personnel who have read and
understand the User Instructions should use MAXAIR Products.
The institution using this product in any application is responsible
for determining the appropriateness of this equipment relative to
regulatory requirements. Bio-Medical Devices Intl, Inc. does not
recommend the appropriate systems for a particular institution or
facility.
Use only MAXAIR Systems/ NIOSH approved compatible
components.
NOT for use in atmospheres immediately dangerous to life or health
(IDLH), and atmospheres containing less than 19.5% oxygen, or
more than 25% oxygen.
Flammability Level I: fabric may burn if exposed to open flame.
Follow current local regulations governing biohazard waste to safely
dispose of single use MAXAIR Products.
If you need more information, contact your BMDI Sales
Representative, or call BMDI customer service at 1-800-443-3842.
!
CAUTION
Before use, thoroughly read MAXAIR CAPR P/N 03521015 User Instruction (received with all CAPR Helmets and
available at www.maxair-systems.com.
!