7 Operating with Digital Control
65
HARS series
OVPSET Command
Sets the overvoltage protection value.
Range
0 to the maximum value of overvoltage protection
Resolution
Refer to
[7-9 Range for Voltage Setting and Crrent Setting].
Unit
V
Format
example
OVPSET 23450
⇒
For V in unit, set 23.45kV.
Initial value
Maximum value of overvoltage protection
Query
In response to
“OVPSET?”, the set value of overvoltage protection.
Return example: OVPSET=23450
Remarks
・
A setting value exceeding the maximum value of overvoltage protection is ignored
as an invalid command.
・
This command is compatible with “Multi-Command”.
Note: Maximum value of overvoltage protection = Rated voltage value × 110%
OCPSET Command
Sets the overcurrent protection value.
Range
0 to the maximum value of overcurrent protection
Resolution
Refer to [7-9 Range for Voltage Setting and Crrent Setting].
Unit
Same as Unit displayed on the front panel.
Format
example
OCPSET 0.02345
⇒
For A in unit, set 23.45mA.
Initial value
Maximum value of overvoltage protection
Query
In response to
“OCPSET?” the set value of overcurrent protection.
Return example: OCPSET=0.02345
Remarks
・
A setting value exceeding the maximum value of overcurrent protection is ignored
as an invalid command.
・
This command is compatible with “Multi-Command”.
Note: Maximum value of overcurrent protection = Rated current value × 110%
Содержание HARS Series
Страница 1: ...Instruction Manual F RA 001 3R3 MODEL HARS series B N 277 9 002 277 9 002 Rev 0 1 ...
Страница 10: ......
Страница 19: ...2 Before Using This Product 9 HARS series 2 6 3 Dimensions a Models 500W 1kW 60kV or lower ...
Страница 20: ...2 Before Using This Product 10 HARS series b Models 500W 1kW 100kV 120kV ...
Страница 21: ...2 Before Using This Product 11 HARS series c Models 2kW 3kW 60kV or lower ...
Страница 22: ...2 Before Using This Product 12 HARS series d Models 2kW 3kW 100kV 120kV ...
Страница 92: ......