background image

 

F-RA-001-5R0 

 
 
 

 

Understand the risk. 
Products generate high voltage and energy.   
An electrician or equivalent knowledge must read the instruction manual to the end and familiarize yourself with 
the proper use and danger of the product.   
If you want the product to be handled by another person, limit it to those with knowledge of electricity, and have 
people who are familiar with correct use and danger of the product conduct education and training before handling 
it, or handle the product under the supervision  of  those  who  are  familiar  with  the  proper  use  and  danger  of  the 
product. Persons, with the exception of the above, must not handle the product. 

 

 

Do not open the product case and panel. 
Do  not  disassemble,  remodel,  or  repair  by  opening  the  product  case  and  panel  yourself.  There  is  a  high  voltage 
area inside, which may possibly cause electric shock and fire.   
Never  remove  the  case  and  panel  of  the  product,  since  the  safety  and  function  of  the  product  will  not  be 
guaranteed. 

 

 

It is not suitable for use in places where there is a possibility of a child being present. 
The product is designed assuming that it will be used in places where there are only adults, such as factories and 
laboratories. Do not use it where there is a possibility of a child being present. 

 

 

Ensure that the product is handled only by a limited number of people. 
Only those who are familiar with the proper use and danger of the product, or those who are educated and trained 
by them should install the product.    Any other person handling the product may cause electric shock, burns and 
injuries. 

 

 

Ground it. 
When the product has a protective ground terminal or GND terminal, connect it to the ground.   
If the input terminal of the product is an inlet type, connect to the power outlet with the ground using the supplied 
power cable or a power cable suitable for the area to be used. If it is not connected to the ground, there is a risk 
of electric shock and fire, which is extremely dangerous. 

 

 

About power supply cable 
If there is a power supply cable attached to the package or product, it cannot be used for other products. In the 
case of a product that does not include a power supply cable, you must purchase an optional power supply cable 
suitable for the operating environment and the area, or you must prepare it yourself. 

 

 

About the connection of the power supply cable 
Connection of the power cable should be done by an electrician or someone who has equivalent knowledge and is 
familiar with the correct usage and danger of the product. Refer to the page about connection of the input cable in 
the  instruction  manual  and  carry  out  work.  For  products  that  have  acquired  safety  standards,  also  refer  to  the 
page that describing overvoltage categories. 

 

 

About input voltage 
For the input voltage of the product, refer to the page describing the input terminal of the product and the input 
voltage in the instruction manual. Do not supply input exceeding the specifications of the product. 

 

 

About the fuse 
The product contains a protective fuse. In the case of a product in which a fuse holder is arranged externally, it is 
possible to replace the fuse. For the replacement method, refer to the instruction manual.   
If the product does not have a fuse holder externally, the user cannot replace the fuse. If the fuse blows out,  do 
not open the case or panel. Please contact our sales representative. 

 

 

Do not modify or damage the cable. 
It may cause electric shock and fire. 

WARNING 

 

 

Содержание AUH Series

Страница 1: ...Rev 0 0 B N 231 9 002 231 9 002 MODEL AUH series Instruction Manual ...

Страница 2: ...g of the symbols and observe each item Depending on the product some symbols may not used 禁止 Indicates content that could cause death or serious injury Indicates content that could cause injury or damage to property Indicates danger warning caution Refer to the instruction manual when using the part with this symbol Indicates where high voltage is handled Touching it may cause electric shock and r...

Страница 3: ... terminal or GND terminal connect it to the ground If the input terminal of the product is an inlet type connect to the power outlet with the ground using the supplied power cable or a power cable suitable for the area to be used If it is not connected to the ground there is a risk of electric shock and fire which is extremely dangerous About power supply cable If there is a power supply cable att...

Страница 4: ...r right hand In case of electric shock when manipulating the product in order to reduce the risk of current flowing through important organs of the body Keep your left hand off from the product and operate with your right hand Do not touch the wiring or load connected to the output terminal or output terminal during operation or immediately after output stop Very high voltage is applied to the ter...

Страница 5: ... port and exhaust port are blocked the specified performance of the product cannot be demonstrated and parts are deteriorate which may cause smoke and fire When work such as welding is conducted near the product disconnect all the wires from the product Do not wipe with chemicals thinner etc or wet cloth There is a danger of electric shock electric leakage burnout from chemicals or water entering ...

Страница 6: ...ram 2 1 2 1 Front panel 2 1 2 2 Display panel detail 2 1 2 3 Rear panel 2 2 2 4 Dimensions 2 3 3 Normal Operation 3 1 3 1 Operation 3 1 3 2 Protection 3 2 3 2 1 Over Voltage Protection O V P 3 2 3 2 2 Over Temperature Protection 3 2 3 2 3 External Voltage Limit Protection 3 2 4 Advanced Operation 4 1 4 1 Introduction 4 1 4 2 Function of TB1 interface connector 4 1 4 3 Selecting internal control IN...

Страница 7: ...nternal INT cut off current control 4 4 4 6 Remote switch ON OFF LS 4 4 4 7 Door switch LD Interlock 4 5 4 8 Output current monitor Output voltage monitor 4 6 4 8 1 Output current monitor L1 4 6 4 8 2 Output voltage monitor L2 4 6 4 9 HV status signal 4 6 5 Option 5 1 5 1 Floating ground LF 5 1 5 2 Slow start LW 5 2 5 3 Overload trip LOc 5 2 5 4 Master Slave LMs 5 3 CONTENTS ...

Страница 8: ...stant current mode 0 to 10Vdc external voltage Voltage Regulation Line 0 02 of maximum voltage for 10 input line change Load 0 02 of maximum voltage for 10 100 load change Current Regulation Line 0 05 of maximum current for 10 input line change Load 0 05 of maximum current 100μA for 10 100 load change Ripple Less than 0 3 p p 1Vrms Less than 0 1 p p 1Vrms optional only L200V3P Stability 0 05 Hr Te...

Страница 9: ...t more than 30cm from rear panel External cabinet shall have the same size of intake hole and exhaust space In case that the power supply is operated in the closed cabinet such arrangement as forced ventilation shall be required so that temperature by exhausted air will not be taken into air intake of power supply Make sure no liquid or metal piece is taken into the vent holes of power supply Do n...

Страница 10: ... ground the power supply main body with the AC line input connector or it is very dangerous The rated input voltage is set when the unit is dispatched from the factory Verify the input voltage indicated on the rear panel and input the correct voltage Unless the input voltage is as shown above due to special specifications verify the input voltage indicated on the rear panel and input the correct v...

Страница 11: ...s dangerous and may cause an electric shock or damage the power supply When load short circuit or discharge is forecasted make the GROUND cable thicker and shorter MUST CAUTION Fig 1 2 output cable load frame ground ground terminal rack mount case Wires should be more than 5SQ less than about 2m not bared load MUST Ground at one point MUST Ground at one point Fig 1 3 load output cable ground termi...

Страница 12: ...n 10 seconds and use another high voltage meter to make sure that the voltage is zero Especially it is dangerous when capacitive load is applied or when long cable is attached 5 Touch the load with your right hand 1 5 Troubleshooting No output voltage 1 Check if the specified voltage is input Power supply input voltage Control voltage 0 10V in external voltage controlling Control current 0 10V in ...

Страница 13: ...ttention 1 This power supply unit need to special insulation oil Please note that the performance cannot be guaranteed when other insulation oil is used 2 This insulation oil uses the one not volatile However please confirm the liquid level has not fallen once every six months 3 Please keep in the cool dark place and keep the lid of the container 4 Please affix the manufacturing type name and orde...

Страница 14: ...de Output current digital meter Output voltage digital meter 2 Exterior view diagram 2 Exterior view diagram Fig 2 1 Front panel 図 2 1 フロントパネル POWER ON OFF switch Output voltage setting dial OUTPUT ON OFF switch Output current setting dial OUTPUT ON LED Ventilation inlets Display panel Fig 2 2 Display panel detail OT LD EXT kV mA CC CV ...

Страница 15: ...anel 2 Exterior view diagram Fig 2 3 Rear panel Output connector Ground terminal LF terminal for LF option Interface connector TB1 P E terminal optional only L200V3P AC input connector Exhaust holes GND TB1 S2 OUTPUT ...

Страница 16: ...2 2 2 4 Dimensions 2 Exterior view diagram 3 ...

Страница 17: ...o ON 1 and the meter lamp is lit for stand by 7 Turn the OUTPUT ON OFF switch to the ON position The red indicator OUTPUT ON LED will light up to indicate that the power supply is ready for operation In case of turning the OUTPUT ON OFF switch to the ON position wait 10 seconds or more and then operate the power supply 8 Turn the output voltage setting dial VOLTAGE to set to an desired output volt...

Страница 18: ...V P Over Voltage Protection 3 2 2 Over Temperature Protection This unit is equipped with the built in circuit to cut off the high voltage output at the set internal temperature or higher After making sure that the OT LED is OFF select the OUTPUT ON OFF switch to OFF to reset 3 2 3 External Voltage Limit Protection This unit is equipped with the external voltage limit protection To protect the powe...

Страница 19: ...of Power Supply For 1 2 3 14 15 6 Spare out of use 7 Door switch Interlock Short HV ON 8 Remote reset for Option LOc Short reset 9 Spare out of use 10 Spare out of use 11 Spare out of use 12 Spare out of use 13 Spare out of use 14 Reference voltage output Reference 10 24V 4mA MAX output 15 Output Voltage Monitor 0 to 10Vdc output 0 to 100 rated output 16 Spare out of use 17 Common GND of Power Sup...

Страница 20: ...NT EXT with control cable for GP IB or RS 232C option is done by sending commands When the GP IB and RS 232C control cable is attached to the rear TB1 the power supply cannot operate alone Precautions in using the open collector 1 Use the open collector according to the following List 4 2 Open collector Output Switch Open collector ON Short VCE Under 0 4V 10mA OFF Open VCE Over 2V Open 5V 5V L EXT...

Страница 21: ...l voltage fluctuation directly influences on the output The external voltage should be stable 4 5 External volume control 4 5 1 External volume voltage control LRv The external volume 5kΩ can control the output voltage within the range of 0 100 4 5 2 External volume current control LRc The external volume 5kΩ can control the output current within the range of 0 100 TB1 Current control Fig 4 3 Exte...

Страница 22: ...n No 2 4 6 Remote switch ON OFF LS The remote switch can select the output to ON OFF The open collector can also do When the output is ON the front panel OUTPUT LED is lit This function is effective regardless of selection of INT EXT Front panel CURRENT dial 14 15 16 17 18 19 20 25 1 2 3 4 5 6 7 8 13 CV ERROR AMP CC ERROR AMP 4mA max 10 24Vdc TB1 CV EXT CC INT Current control 10 24Vdc Voltage cont...

Страница 23: ... ON The open collector can be also used for the purpose Open the switch 3 and the front panel LD is ON Once select the OUTPUT ON OFF switch to OFF and next select to ON again to reset When LD is used open the switch S2 This function is effective regardless of selection of INT EXT When the door switch is not used Short circuit the switch S2 and the LD will not function Keep pressing the upper squar...

Страница 24: ...voltage The output impedance is 1kΩ The monitor output polarity is positive regardless of the high voltage output polarity 4 9 HV status signal When the OUTPUT ON OFF switch is ON the internally built in relay contact is ON Caution When the contact is open the voltage is 30Vdc MAX and the permissible current is 100mA MAX TB1 14 15 16 17 18 19 1 2 3 4 5 6 VOLTAGE MONITOR AMP Z out 1kΩ Fig 4 7 Outpu...

Страница 25: ...erminal See Fig 5 2 When this function is not used attach the short circuit fitting between the CASE terminal and the 0V terminal to connect See Fig 5 2 When this function is used the device connected to TB1 should be floating Fig 5 1 Main body Load and Ground of the unit equipped with this function Fig 5 2 Example of the unit equipped with this function CASE 0V A With this function Without this f...

Страница 26: ...the maximum rated voltage can be attained in 10 seconds approximately 5 3 Overload trip LOc When the output current exceeds value set by the CURRENT dial or external voltage the output is cut off During cutting off the CC LED is ON and the OUTPUT LED is OFF Once select the OUTPUT ON OFF switch to OFF and next select to ON again to reset The external switch can reset the cut off condition at output...

Страница 27: ...d slave unit by connection of AUH LMs cable The wire connection of attached AUH LMs cable is shown above and standard local control of output voltage and current is available setting dials on front panel of master unit When monitoring the output or status connect measurement equipment to master unit Switch the of external Change 2 connections of connectors at master unit side shown below Refer to ...

Страница 28: ......

Отзывы: