If the caliper finds a negative diameter reading, an error notification is shown on the
display: "Calibrating Error" (the measuring jaws were not together when switching on)
1. Switch the caliper off.
2. Bring the measuring arms together.
3. Switch the caliper on.
4. Check the measuring accuracy with the calibration plate.
2.4 CONNECTIONS
The cable set delivered with the caliper can be connected directly from docking
station
to PC’s serial port. One end of the cable must be connected to 12 VDC fuse
box and protected with max. 4A fuse.
Excaliper can also be connected to PC via 9
– pin Sub D – connector. Connection is
serial RS 232. The serial cable delivered with the caliper can be connected directly to
PC’s serial port. Connect the cable first to the PC and then to the caliper. It is
recommended that the caliper is switched off while connecting the cables.