Rad-97
Chapter 4: Operation
www.masimo.com
69
Masimo
Options
Description
Alarm
Priority
Factory
Default
Settings
User Configurable
Settings
Respiratory
Pause
The duration of time that
triggers an alarm if no
breaths are detected.
NA
30 seconds 15, 20, 25, 30, 35, or 40
seconds
Alarm Delay When a High or Low alarm
condition occurs, this
feature delays the audible
part of an alarm.
NA
30 seconds 0, 10, 15, 30, or 60
seconds
Additional Settings for RRa
From the
Additional Settings
screen, change any of the following options:
Options
Description
Factory
Default
Settings
User Configurable
Settings
Averaging
Time
The length of time over which the
system calculates the average of all
data points.
Slow
Trending, No
Averaging, Fast,
Medium, or Slow
Freshness
The duration of time that, during
interference, the system displays the
last valid reading.
5 minutes
0, 1, 5, 10, or 15
minutes
RRp Settings
RRp is currently not available in the U.S.A. and territories relying on FDA market
clearance.
When using a pulse oximetry or pulse CO-Oximetry sensor with Rad-97, respiration rate can
be determined by the plethysmographic waveform (RRp). This method measures a patient's
respiratory rate based on plethysmographic amplitude changes corresponding to the
respiratory cycle. When using a pulse oximetry or pulse CO-Oximetry sensor, RRp alarms and
RRp settings are active and the
Main Screen
labels respiratory rate as
RRp
, as shown below.
Содержание Rad-97
Страница 1: ...Operator s Manual Rad 97 Pulse CO Oximeter...
Страница 2: ......
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 36: ......
Страница 104: ......
Страница 110: ......
Страница 118: ......
Страница 124: ......
Страница 146: ......
Страница 168: ......
Страница 178: ......
Страница 186: ......
Страница 187: ......
Страница 188: ...www masimo com 38281 LAB 9275B 0917...