Chapter 6: Alerts and Messages
www.masimo.com
64
Masimo
Acknowledging Alerts
Alerts are conveyed visibly.
To acknowledge alerts:
1.
Touch the
Alert Acknowledgment
icon
.
2.
Visual alerts are suspended.
3.
Explanation of the alert remains in the Status Bar.
Messages
The following section lists common Rad-67 messages, potential causes, and next steps.
Message
Potential Causes
Next Steps
(Pulse CO-Ox)
Replace Cable
or
Please replace
cable
Incompatible cable
Unrecognized cable
Defective cable
Cable life expired
Replace the patient cable.
(Pulse CO-Ox)
Cable Near
Expiration
Cable life near expiration
Replace the patient cable.
(Pulse CO-Ox)
Replace
Sensor
or
Please replace
sensor
Incompatible sensor
Unrecognized sensor
Defective sensor
Sensor life expired
Replace the sensor.
(Pulse CO-Ox)
Sensor
Initializing
Device is initializing the
sensor.
Allow time for parameter reading
to stabilize.
(Pulse CO-Ox)
Pulse Search
Device is searching for
pulse.
Allow time for parameter reading
to stabilize.
Содержание Rad-67 Pulse CO-Oximeter
Страница 1: ...Operator s Manual Rad 67 Pulse CO Oximeter Spot Check...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 12: ......
Страница 20: ......
Страница 58: ......
Страница 76: ......
Страница 94: ......
Страница 100: ......
Страница 104: ......
Страница 105: ......
Страница 106: ...www masimo com 300362 LAB 10262A 0419...