Martini Instruments MI 415 Скачать руководство пользователя страница 1

USER MANUAL

MI 415

Turbidity Meter

Dear Customer,
Thank you for choosing a Martini product.
This manual will provide you with the neces-
sary information for the correct use of the
instrument. Please read it carefully before
using the meter.
This instrument is in compliance with 
directives

.

Range

0.00 to 50.00 FNU
50 to 1000 FNU

Resolution

0.01 FNU and 1 FNU

Precision

±0.5FNU or ±5% of read-

ing, whichever is greater

Light Source

High emission infrared LED

Light Detector

Silicon Photocell

Method

Detection of scattered light

Environment

0 to 50°C (32 to 122°F);

100% RH max.

Battery Type

1 x 9 volt

Auto-Shut off

After 5' of non-use

Dimensions

192 x 104 x 52 mm

(7.5 x 4.1 x 2”)

Weight

380 g

SPECIFICATIONS:

ISTMI415  07/05

MEASUREMENT PROCEDURE:

1• Turn the meter on by pressing

ON/OFF.

2• When the LCD displays “- - -”, the

meter is ready.

3• Fill a clean, dry cuvet with sample

up to the mark.

4•  Replace the cap.

5• Clean the cuvet with a lint-

free cloth to remove any fin-
gerprints, dirt or waterspots.

6• Place the cuvet into the

holder and ensure that the
notch on the cap is posi-
tioned securely into the
groove.

7• Press READ/   and “SIP” will blink during

measurement.

8• The instrument directly displays turbidity in

FNU units.

CALIBRATION PROCEDURE:

1• Turn the meter on by pressing ON/OFF.

When the LCD displays “- - -”, the meter is
ready.

2• Press CAL key and “CAL” will blink on the

display.

3• Press CAL key again. The date of calibra-

tion in MM.DD format appears on the LCD.

4• To toggle between month and day

press DATE/    key.

5• To change the date press READ/ 

key.

6• To confirm the date values press

CAL key.

A blinking ZERO message will appear on
the LCD. Insert the 0 FNU standard and
press CAL key. “SIP” will blink during mea-
surement.